Ultra-high contrast MRI of the brain and spinal cord using directly acquired and synthetic BipoLAr Inversion Recovery (BLAIR) images
Introduction
Ultra-high contrast (UHC) magnetic resonance imaging (MRI) is a term used to describe magnetic resonance (MR) imaging that shows abnormalities with high contrast when little or no abnormality is seen using common conventional state-of-the-art MRI sequences. It is achieved without the use of increased static or gradient magnetic fields. For soft tissues, X-ray computed tomography (CT) can be regarded as high contrast, conventional state-of-the-art MRI as very high contrast and MRI which shows abnormalities with obvious contrast when they are not seen with conventional MRI as UHC (1).
A common mechanism for achieving UHC MRI is synergistic contrast in which changes in a tissue property such as T1 or T2 are used twice or more in the same sequence to increase contrast rather than just using changes in a single property once which is the case with most conventional sequences (2). An example is the divided subtracted inversion recovery (dSIR) sequence in which the signals from two inversion recovery (IR) sequences with different inversion times (TIs) are subtracted to increase the contrast produced by small changes in T1. This contrast is increased further by division of the subtraction by the sum of the signals from the two sequences (3).
Another way of achieving UHC MRI is to use a high-quality single tissue property map such as a T1 or T2 map and retrospectively apply a shaped filter to the map. The filter can increase contrast in one or more domains of tissue property values and maintain anatomical detail in the rest of the image. This is unlike narrow windowing of images which, when used to increase contrast, is accompanied by saturation of the display gray scale at the upper and lower signal thresholds. As a result, narrow windowing leads to loss of anatomical detail and limits the degree to which contrast can be increased by windowing.
The principal clinical focus of UHC MRI to date has been on small changes in tissue properties from normal. When normal and/or abnormal structures are already seen with high contrast using conventional sequences, there is no particular clinical gain in demonstrating them with even greater contrast. As a result, UHC MRI has been directed at normal or near normal appearing tissues seen with conventional state-of-the-art images where there is little or no apparent lesion contrast with the aim of demonstrating abnormalities with high contrast.
Small changes in T1 may produce obvious abnormalities with the dSIR sequence as illustrated in a case of mild traumatic brain injury (mTBI) in Figure 1. In an age matched normal control no abnormality is seen on the T2-FLAIR image (Figure 1A) or the corresponding dSIR image (Figure 1B). The normal dSIR image shows normal white matter in the cerebral hemispheres as low signal (mid gray) or very low signal (dark) in Figure 1B. In the patient, a 24-year-old male examined after mTBI, the T2-FLAIR image (Figure 1C) is normal, but the corresponding dSIR image shows abnormal high signal (light appearance) in the entire white matter (white arrows) (Figure 1D). There is a night and day difference between the normal white matter in Figure 1B and the abnormal white matter in Figure 1D. No abnormality is seen on the corresponding T2-FLAIR images. High signal, well-defined boundaries are seen between normal white and gray matter on the normal dSIR image in Figure 1B. These boundaries are less obvious in the abnormal (Figure 1D) because of the high signal present in abnormal white matter.
High contrast abnormalities in white matter have been seen using dSIR T1-BipoLAr Inversion Recovery (T1-BLAIR) images of the type illustrated in Figure 1 in areas where little or no abnormality has been seen on T2-FLAIR or T2-wSE images in cases of mTBI (4), multiple sclerosis (MS) (5,6), methamphetamine substance use disorder (6) and Grinker’s myelinopathy (7).
The purpose of this educational review is to describe the basic physics underlying four groups of BLAIR images and illustrate their use in normal human subjects and patients.
Basic physics
Tissue property filters (TP-filters) and the IR sequence
The first part of this paper is a condensed and updated version of work published previously (1-3). It is included to make the paper self-contained and not require concurrent reference to other papers to understand the contrast seen on BLAIR images included in this paper.
Instead of illustrating image contrast in the usual way by plotting longitudinal and then transverse magnetisation (MZ and then MXY) against time using the Bloch equations as in Figure 2, it is possible to use plots of signals produced by the exponential recovery of longitudinal magnetisation and exponential decay of transverse magnetisation against T1 and T2 respectively (rather than against time). These plots are TP-filters; the tissue properties may be mobile proton density (ρm), T1, T2, T2*, D*, etc. (8,9). (In this paper the term tissue is used to include fluids unless otherwise specified.)
For TR >> T1, the T1 dependent part of the IR sequence has the signal equation:
where ST1 is the signal from the T1-filter, TI in the inversion time and T1 is the longitudinal relaxation time. The T1-filter of the IR sequence (with a constant value of TI) is shown in phase-sensitive or signed form in Figure 3A where it is a monotonic low pass filter, and in magnitude form in Figure 3B where it is a negative unipolar T1-filter. The maximum size of the slopes of these two filters is shown with red lines in Figure 3A,3B. The magnitude forms of the IR filter shown in Eq. [1] are used for and for in the subsequent text unless otherwise specified.
When TI is increased, the magnitude IR unipolar T1-filter shifts to the right as shown in Figure 4. Figure 4A (left) shows the IR T1-filter with a short TIs (e.g., the short TI IR or STIR sequence) for brain where gray matter (G) has a higher signal than white matter (W). The slope of the T1-filter between W and G is moderately positive (red line).
When the TI is increased to an intermediate TI, TIi, as in Figure 4B (centre), the T1-filter is shifted to the right. W and G are fixed in the same position on the X axis, and W now has a higher signal than G. The slope of the T1-filter between W and G is strongly negative (red line).
When the TI is increased further to a long TI, TIl, the T1-filter is displaced further to the right, as in Figure 4C (right) where W has a slightly higher signal than G, and the slope of the T1-filter between them is mildly negative (red line).
Using the small change approximation of differential calculus, the contrast (difference in signal) produced by each T1-filter from the increase in T1 from W to G (horizontal green arrows in Figure 4A-4C) is the size of this increase multiplied by the slopes of the respective T1-filters (i.e., by their first derivatives, the red lines). This contrast is shown by the vertical blue arrows in Figure 4 and is moderately positive in Figure 4A, highly negative in Figure 4B and mildly negative in Figure 4C. In each example, the slopes of the T1-filters show the contribution of the sequence to contrast.
Subtracted IR (SIR) and dSIR bipolar filters
Figure 5A shows a dSIR bipolar T1-filter in which an SIR bipolar T1-filter is divided by an added IR (AIR) T1-filter i.e., to give a dSIR filter (blue curve). The signal (SdSIR) of the dSIR filter is given by:
where STIs is the signal from a shorter TI filter and STIi is the signal from an intermediate TIi filter.
The resulting dSIR bipolar T1-filter (Figure 5A) shows a negative pole and a positive pole (blue curve). It has a middle domain (mD) between the two poles (vertical dashed lines) which shows a highly positive nearly linear slope (red line). This is about ten times greater than the slope (red line) of the IR unipolar T1-filter shown in Figure 5A (pink curve). The dSIR filter has maximum and minimum values of ±1. The contributions from ρm and T2 in conventional IR sequences cancel out with dSIR sequences which are therefore T1 maps. As a result, the whole sequence can be accurately represented by its bipolar T1-filters.
Figure 5B compares the contrast produced by the IR unipolar T1-filter (pink) to that produced by the dSIR bipolar T1-filter shown in Figure 5A (blue). The increase in T1 (horizontal positive green arrow, ∆T1) is multiplied by the slopes of the respective IR and dSIR filters (red lines) to produce the differences in signal ∆S, or contrast generated by the two filters. These are shown by the vertical pink and blue arrows on the right. For the same increase in T1 (∆T1), the dSIR bipolar T1-filter produces about ten times greater contrast than the IR unipolar T1-filter. The IR T1-filter is that of a conventional TI IR sequence such as MP-RAGE (magnetisation prepared-rapid acquisition gradient echo).
To produce the large increase in contrast shown in Figure 5B, the dSIR sequence is typically targeted at the small increase in the T1 of normal tissue produced by disease (shown by the horizontal green arrow) and this is positioned within the steeply sloping mD of the dSIR bipolar T1-filter. This is done by choosing appropriate values of TIs and TIi. The target is often small increases in the T1 of white matter. High contrast can also be produced by difference or change in T1 within the lower part of the hD and the upper part of the lD where the dSIR bipolar T1-filter has relatively steep slopes. More detail is available in reference (3).
Contrast at tissue boundaries
MRI tissue boundaries take different forms such as a gradual change in signal from one tissue to another as well as sharply defined high and low signal boundaries between tissues. In the boundary region between two pure tissues (such as between white and gray matter) the T1s of voxels which contain mixtures of the two tissues typically span the range of T1 values between those of the two pure tissues. If a narrow mD dSIR bipolar T1-filter [e.g., with TIs (TI short) nulling normal white matter, and TIi (TI intermediate) longer than TIs but less than that needed to null gray matter] is used there are mixtures of white and gray matter in voxels in the boundary region between the two tissues which have intermediate T1 values (T1W,G) that correspond with the peak of the bipolar T1-filter as shown in Figure 6. This produces a high signal boundary between white matter and gray matter. The dSIR bipolar T1-filter shows much higher contrast than that produced between white matter and gray matter by a conventional white matter nulled IR sequence such as MP-RAGE (Figure 6). The mechanism shown in Figure 6 produces the high signal SW,G at the boundary between normal white and normal gray matter shown in Figure 1B. High signal boundaries are also seen at the junction between cerebrospinal fluid (CSF) and white matter at ventricular boundaries, and can be seen at the boundary between gray matter and CSF with wide mD sequences (3).
T1 maps and qualitative—quantitative MRI
To better understand the dSIR bipolar T1-filter, a linear equation of the form y = mx + c can be used to approximate the filter in its mD. The equation is produced by fitting a straight line between the first and last points of the mD (i.e., first point and y =−1, and last point and y = +1). In the mD, the signal of the dSIR sequence SdSIR, is given by:
where ∆TI = TIi − TIs (i.e., second TI minus first TI) which is positive, and ΣTI = TIs + TIi which is also positive. Note that because ∆TI is positive, the slope is positive. The offset is negative.
The expression in Eq. [2] illustrates four key features of the dSIR bipolar T1-filter, firstly, the near linear change in signal (i.e., SdSIR) with T1 in the mD, secondly, the filter has a slope in the mD equal to , thirdly the filter shows high sensitivity to small changes in T1 when the size of ∆TI is small. When ∆T1 is small, the size of ∆TI can be decreased to match it and so scale up the sensitivity of the filter. The reduction in ∆TI increases the steepness of the T1-filter in the mD and thus the amplification of contrast produced from ∆T1. This compensates for the small value of ∆T1 until the sequence becomes signal-to-noise ratio (SNR) and/or artefact limited. This is not the case with conventional IR sequences where, if ∆T1 is decreased, contrast decreases.
Fourthly, Eq. [3] can be used to calculate T1 values in the mD so that:
The linear approximation is only valid in the mD. Also, it is assumed that repetition time (TR) is long compared to tissue T1 values. An equivalent expression that allows for incomplete recovery of longitudinal magnetisation during TR can be formulated.
Outside of the mD in the lowest domain (lD) and highest domain (hD), the magnitude of the dSIR bipolar T1-filter signal decreases towards zero at minimum and maximum values of T1. T1s in the mD occupy the full range of the display gray scale from black to white. T1s from the shortest and longest T1s in the lD and hD respectively only occupy parts of the display gray scale.
Magnetisation transfer (MT) (see later) may have a substantial effect on T1 values which can be regarded as observed T1s reflecting the process of signal acquisition as well as T1 as an intrinsic property of tissue.
Logarithmic then subtracted inversion recovery (lSIR) sequences
Another bipolar T1-filter is the natural lSIR filter (9). The signal of this filter (SlSIR) is half of the subtraction: ln short STIs T1-filter minus ln intermediate STIi T1-filter. The ln STIs and ln STIi negative unipolar T1-filters are shown in Figure 7 and have sharper negative poles than those of unipolar STIs and STIi T1-filters shown in Figure 5A,5B.
Figure 8 shows a dSIR bipolar T1-filter in blue and the corresponding lSIR bipolar T1-filter with the same TIs in orange. The two filters appear similar in the lowest region of the lD and in the highest region of the hD as well as in the central region of the mD. The slope of the dSIR filter is essentially constant and equal to . The magnitude of the slope of the lSIR filter is the same that of the dSIR filter at the centre of the mD but greater around the lower and higher nulling TI values. The lSIR filter asymptotically approaches negative and positive infinity at the two corresponding T1s. The lSIR filter increases its contrast amplification as the change in T1 from normal becomes smaller. Like the dSIR filter, the lSIR filter essentially eliminates the ρm and T2 dependence of the full IR sequence and so can be used to model the full behaviour of the sequence.
The lSIR filter is an inverse hyperbolic tangent of the dSIR filter in the mD (SlSIR = atanh SdSIR), and an inverse hyperbolic cotangent of the dSIR filter in the lD and hD.
From a practical point of view, the lSIR T1-filter has 2–3 times the slope of the dSIR T1-filter for very small differences or changes in T1 around the null points and thus has 20–30 times greater contrast than conventional IR sequences for the same very small differences or changes in T1. The maximum and minimum values of the lSIR T1-filter are usually shown as ±2–3 (rather than ± infinity). This compares with values of ±1 for the dSIR T1-filter. The low and high signal boundaries of the lSIR T1-filter are narrower than those of the corresponding dSIR filter. The lSIR T1-filter shows a selective increase in slope, or sharpening, close to the null points.
Composite (c) bipolar filters (T1 as well as T2, T2*, D*, χ and/or MT)
It is possible to introduce an attenuation filter SOF (other filter) and apply it to one of the two IR filters used for dSIR bipolar T1-filter imaging shown in Figure 5. SOF may equal e–∆TE/T2, e–∆TE/T2* or e–∆bD* which introduce T2, T2* and D* dependence respectively. This provides additional contrast to the T1 contrast of the dSIR bipolar T1-filter. When ∆TE equals zero, or ∆b equals zero, SOF =1. If SOF is set to 0.5 and 0.25, the curves shown in Figure 9A,9B result when using a narrow mD dSIR sequence (TIs =350 and 500 ms). The nulling times are the same for all values of SOF.
In Figure 9A attenuation of the first IR filter (with TIs) is shown with SOF set at 1 (blue), 0.5 (orange) and 0.25 (yellow). The yellow and orange curves are wider around the outside of the first (negative) pole at the first TIs and are narrower and inside the second (positive) pole at TIi. This results in a broadening and loss of contrast and lower signal white matter around the first negative pole as well as increased contrast and narrower boundaries between white and gray matter around the positive pole. SOF can be set to negative values to reverse the sign of the relevant contrast and make it synergistic with the T1 contrast.
In Figure 9B, attenuation of the second IR filter (with TIi) is shown. As SOF is decreased from 1.0 to 0.5 and 0.25, the filter narrows around the first (negative) pole and widens around the second (positive) pole. The negative contrast produced for the same difference in T1 is increased around the first negative pole. Composite (c) forms of the dSIR and lSIR bipolar T1-filters can be designated as cdSIR, clSIR etc.
Susceptibility and MT can also be used to attenuate signals and produce supplementary contrast. The combination of T1 and T2* contrast may be of value in demonstrating the central vein sign and paramagnetic rim sign in MS as well as late sequelae of haemorrhage and fMRI in combination with perfusion.
Synthetic dSIR and lSIR bipolar T1-filter images
Synthesizing narrower mD images from wider mD dSIR and lSIR bipolar T1-filter images
It is possible to synthetically create narrower mD dSIR images from wider mD dSIR images using Eq. [1] and the relationships illustrated in Figure 5A. The TIs of the narrower mD images are within those of the wider mD when using magnitude forms of the IR sequences. MT effects may need to be included.
Synthesizing dSIR and lSIR bipolar T1-filter images from T1 maps
dSIR and lSIR bipolar T1-filter images with particular values of TI can also be synthesised from T1 maps using Eq. [1] and the relationships shown in Figures 5A,7,8. T1 and other tissue property maps (T2, T2*, D*) have linear TP-filters with maximum signal values at the high end of the tissue property X axis. As a result, CSF has high signals on the maps and this can dominate their appearance. The problem can be avoided when appropriate bipolar T1-filters are applied to T1 maps to create bipolar T1 images in which CSF signal is reduced. The equivalent mapping using magnetisation prepared 2 rapid acquisition gradient echo (MP2RAGE) acquisitions has also been described previously. While in principle the same contrast is obtained, there are wide discrepancies in the T1s measured by different techniques suggesting T1 is not independent of the method of acquisition with MT a major contributor to the variance. This variance may mask small white matter abnormalities seen with general purpose T1 mapping.
Synthesizing T2-, T2*-, D* and χ-bipolar filter images using purpose designed bipolar filters together with T2, T2*, D* and χ maps
It is also possible to create bipolar TP-filter images from tissue properties other than T1 using purpose designed synthetic bipolar TP-filters. These filters are typically based on the appearance of the dSIR bipolar T1-filter shown in Figure 5A,5B. They can have a linear function in their mD and a scaled reciprocal function (e.g., ) or similar function in their hDs and lDs. These functions join the linear function in the mD at signal values of −1 and +1, respectively. Synthetic bipolar TP-filters can then be used with T2, T2*, D* and χ maps to create bipolar T2-, T2*-, D*-and χ-filter images.
The lSIR inverse hyperbolic tangent and cotangent functions can also be used in the same way to create synthetic lSIR bipolar TP-filter images from tissue property maps.
Directly acquired bipolar T1-filter images and synthetic bipolar T1-, T2-, T2*-, D*-, and χ-images with different TPs can be multiplied together to produce synergistic bipolar multi TP-filter images. The signs of the slopes of the bipolar TP-filters (negative or positive) can be matched with the signs of the changes in tissue properties (negative or positive) to ensure the contrast produced by the different tissue properties is synergistic (2).
Window levels and widths of displayed images
Conventional changes in window level and width of MR images
When an image is displayed, the window level sets the brightness (signal level) and the window width sets the contrast (signal difference). Changes in the window level and width of images change their signals on the gray scale display from S to SD and can be described using a high pass filter as seen in Figure 10A. This plots the display signal SD against the image signal S. This is a signal (S) filter not a TP-filter. Contrast (the difference in signal between the SDs of P and Q) may be increased by narrowing the window width as shown in Figure 10B so that P and Q are now respectively shown at the lower and upper ends of the display scale. With narrow windowing, signals with values at the upper end of the X axis from Q and beyond are all shown with the same high value of SD. As a result, there is no contrast between these voxels and they coalesce into blocks. This leads to a loss of anatomical coherence. The same also happens at the lower end of the S scale where values of S less than P have the same signal in Figure 10B and coalesce. This saturation of signals establishes a practical limit to how tightly images can usefully be windowed (i.e., how narrow the window width can usefully be made) and thus how much the contrast of the displayed image can be increased before the loss of anatomical coherence leads to a lack of credibility of the images. This is in addition to limits imposed by SNR considerations and image artefact levels.
dSIR images
dSIR images can decrease ∆TI and so increase contrast. Maximum and minimum values are shown as high and low signal boundaries as in Figure 10C, T1 values lower than the lower nulling T1 value and greater than the greater nulling T1 value do not appear on the same plateaux as with conventional narrow windowing in Figure 10B, but vary in signal because of the sloping regions in the lDs and hDs of the dSIR bipolar T1-filter (Figure 10C). Consequently, contrast is maintained and anatomical detail is preserved in all three Domains when ∆TI is decreased. As a result, contrast can be increased by decreasing ∆TI until it becomes SNR and/or artefact limited.
Methods
With approval from the New Zealand Health and Disability Ethics Committee 20/CEN/246 (26 June 2020), 21/CEN/246 (1 May 2021); 20/NTB/14 (21 May 2020); UAHPEC 018466 (20 Dec 2019); and the Northern B Health and Disability Ethics Committee 2022 EXP 13106 (18 Nov 2022) as well as informed consent from each subject, MR scans were performed on four adult normal controls, seven adult patients with mTBI, MS, methamphetamine substance use disorder, Grinker’s myelinopathy, Parkinson’s disease and white matter in a patient with a glioma. The research was conducted in accordance with the Declaration of Helsinki. 3T scanners (General Electric Healthcare Premier, Chicago, Illinois, USA and Siemens Healthineers Skyra, Malvern, Pennysylvania, USA) were used. 2D IR fast spin echo (FSE) sequences were performed with a TIs chosen to null the shortest T1 normal white matter (or slightly shorter) and a longer TIi chosen to produce narrow mD dSIR images targeted at small increases in the T1 of white matter from normal as illustrated in Figure 5B. 3D wide mD dSIR images with a short TIs and a long TIl were also obtained (5). Longer TE 2D IR images were acquired to produce cdSIR images. Positionally matched 2D and 3D T2-FLAIR images were obtained as well as susceptibility weighted filtered images as described in Table 1.
Table 1
| No. | Sequence | TR (ms) | TI (ms) | TE (ms) | Matrix size and voxel size (mm) | Number of slices | Slice thickness (mm) |
|---|---|---|---|---|---|---|---|
| 1* | 2D IR FSE (for white nulling matter) | 6,000–9,000 | 350 | 7 | 256×224; 0.9×1.0; Z512 0.4×0.4 | 26 | 4 |
| 2* | 2D IR FSE (used with No. 1 for narrow mD dSIR) | 6,000–9,000 | 500 | 7 | 256×224; 0.9×1.0; Z512 0.4×0.4 | 26 | 4 |
| 3* | 2D IR FSE (for longer T1 nulling) | 6,000–9,000 | 600 | 7 | 256×224; 0.9×1.0; Z512 0.4×0.4 | 26 | 4 |
| 4* | 2D IR FSE (used with No. 1 for wide mD dSIR) | 6,000–9,000 | 800 | 7 | 256×224; 0.9×1.0; Z512 0.4×0.4 | 26 | 4 |
| 5 | 3D BRAVO (for white matter nulling) | 2,000 | 400 | 6 | 256×256; 0.8×0.8; Z512 | 220 | 0.8 |
| 6 | 3D BRAVO (used with No. 5 for wide mD dSIR) | 2,000 | 800 | 6 | 256×256; 0.8×0.8; Z512 | 220 | 0.8 |
| 7* | 2D T2-FLAIR | 6,300 | 1,851 | 102 | 320×240; 0.7×0.7; Z512 0.4×0.4 | 26 | 4 |
| 8 | 3D T2-FLAIR | 6,300 | 1,850 | 102 | 256×256; 0.8×0.8; Z512 0.6×0.6 | 252 | 0.8 |
| 9 | 3D susceptibility weighted | 40 | – | 32 | 300×300; 0.8×0.8; Z512 | 110 | 2 |
*, positionally matched for slice position, slice thickness, slice number, field of view and interpolated matrix size. 2D, two-dimensional; 3D, three-dimensional; BRAVO, BRAin VOlume; dSIR, divided subtracted inversion recovery; FLAIR, fluid attenuated inversion recovery; FSE, fast spin echo; IR, inversion recovery; mD, middle domain; TE, echo time; TI, inversion time; TR, repetition time; Z, zipped.
Illustrative cases
mTBI
Conventional MRI images usually show little or no change in mTBI but narrow mD dSIR images frequently show extensive high signal areas in white matter (the whiteout sign) (Figure 1). This may resolve within as little as two days or persist over years. It is generally attributed to neuroinflammation but oedema, demyelination and degeneration may also have a role.
In addition to the whiteout sign, there may be loss of contrast in the gray matter of the thalamus due to an increase in T1. This is manifest as a low contrast appearance between the medial and lateral thalamus (Figure 11A) (arrows) (a grayout sign). After remission normal high gray white matter contrast is restored (Figure 11B) (arrows). Contrast between the lateral thalamus (arrows) and the posterior limb of the internal capsule (PLIC) is low in (Figure 11A) but very high in (Figure 11B). The low contrast between the thalamus and the PLIC in (A) results from a reduction in signal in the gray matter of the lateral thalamus due to an increase in its T1 in the hD as well as an increase in signal in the white matter of the PLIC due to an increase in its T1 in the mD. This is reversed in (Figure 11B) where the normal high contrast boundary is restored.
Figure 12A shows a normal control with obvious contrast between the heads of the caudate nuclei and the adjacent CSF as well as between the cortex and CSF. Figure 12B shows a patient with mTBI who has a whiteout sign. There are grayout signs in the thalamus and putamen.
There may also be gray matter changes elsewhere in mTBI. The normal red nuclei have relatively short T1s resulting in their normal signal being mid-gray in the mD (arrows) (Figure 13A). They show a higher signal due to increase in T1 in a case of mTBI (arrows) (Figure 13B).
MS
Focal lesions, as well as patchy white matter changes, may be seen in areas that appear normal on T2-wFSE images in patients in remission. In a patient examined during a relapse, one lesion is seen on the T2-FLAIR image in Figure 14A (long arrow) is also seen in Figure 14B (long arrow). There are an additional six lesions (small arrows) in Figure 14B. Five of the lesions in Figure 14B have high signal boundaries. This can be due to a large increase in T1 in the lesion beyond the peak of the bipolar T1-filter or a decrease in the abnormal T1 in the lesion due to the presence of paramagnetic free iron. Two of the lesions with high signal boundaries shown in Figure 14B have paramagnetic rim signs on the filtered susceptibility weighted images (arrows, Figure 14C). The high signal boundaries seen on some MS lesions may be a sign of disease activity.
In addition, there are widespread abnormal areas in white matter which are only seen on the dSIR images (Figure 14B). These changes are typically bilateral, symmetrical and have an increased signal. They have the features of a whiteout sign and to date have only been seen in patients with MS during a relapse.
Composite T1 and T2 cdSIR images may show higher contrast than corresponding dSIR images (Figure 15A,15B).
In the spinal cord in another patient diagnosed with MS, the T2-FLAIR image only shows an ill-defined smudge (arrow) (Figure 16A). The corresponding wide mD dSIR image (Figure 16B) shows an extensive, well defined, high contrast abnormality (arrows). Another lesion is seen in the medulla on the dSIR image (arrow) but not on the T2-FLAIR image.
Methamphetamine substance use disorder
Patients with methamphetamine substance use disorder may show a heterogenous pattern of white matter abnormalities with dSIR sequences but they may also show clearly defined whiteout signs and these can remit with continued abstinence (Figure 17) (compare with normal appearance in Figure 1B). The situation may be confounded in the case shown by a previous mTBI.
Delayed post-hypoxic leukoencephalopathy (Grinker’s myelinopathy)
This is thought to be a rare syndrome in which classically patients make an initial recovery after a hypoxic and/or ischaemic episode, and then deteriorate with severe neurological signs such as Parkinsonism and akinetic mutism. Conventional MRI shows widespread abnormalities in the white matter of the cerebral hemispheres. In post-hypoxic patients, it is possible to see extensive abnormalities in white matter with dSIR sequences where no abnormalities are seen with conventional sequences (7). The patients typically have cognitive symptoms of less severity than those included in classical descriptions. This less severe form of Grinker’s myelinopathy may be much more common than the classical form but not be recognised radiologically using conventional MRI sequences.
Parkinson’s disease
Parkinson’s disease may show more obvious bilaminar signs in the cortex than in age matched controls (i.e., an increase in signal in the outer layer of the cortex) with narrow mD dSIR sequences (white arrows) (Figure 18). The inner higher signal is from the boundary between white matter and gray matter and the next layer outwards is lower signal in the inner cortex. The next outer layer beyond this is the higher signal layer in the outer cortex.
Bubble signs in which circular areas of similar size are seen in the gray matter in the basal ganglia and thalamus can also be seen (e.g., black arrows) (Figure 18).
White matter associated with cerebral tumours
In cases of cerebral tumours, extensive white matter changes have been seen after chemotherapy in areas that appeared normal on T2-FLAIR images (Figure 19A,19B). When typical radiation or chemotherapy changes have been seen on T2-FLAIR images, additional changes have been observed in the form of whiteout signs.
Normal control shown with dSIR and lSIR images
dSIR and lSIR images (Figure 20A,20B as well as Figure 20C,20D) are compared in a 46-year-old male normal control. White matter-gray matter boundaries and the bilaminar cortex sign (arrows) are seen with higher contrast and higher spatial resolution on the lSIR images. Bubble signs are also better seen in the thalamus and putamen on the lSIR images. A small decrease in T1 through the sharp peak of the lSIR bipolar T1-filter results in finer boundaries and narrower margin bubble signs with the lSIR bipolar T1-filter compared to the dSIR bipolar T1-filter.
Discussion
UHC MRI using bipolar filters
UHC MRI is a term used to describe MRI techniques which show much greater contrast than conventional state-of-the-art MRI images while having similar spatial resolutions and comparable acquisition times. This is achieved with no increase in static or gradient magnetic field. UHC MRI using bipolar filters employs sequences which exist on most MRI machines together with minimal image processing. It can be implemented in a very short time at very little cost.
UHC MRI is distinguished from ultra-high field MRI which uses static magnetic field strengths of 7T or higher to improve image signal-to-noise ratio. This improvement may be used to increase contrast, spatial resolution and/or the speed of MRI scans (9-13).
UHC MRI is also distinguished from ultra-high spatial resolution MRI where typically increased gradient strength is used to improve spatial resolution (14,15).
The four groups of BLAIR sequences discussed in this paper are summarised in Table 2 including some of their variants and the dominant source(s) of contrast for each sequence. The sequence signal equations and other relevant functions are included in Table 3. The Tables also provide information for instituting synthetic imaging although relevant MT effects may need to be added.
Table 2
| Group | BLAIR sequence | Reverse subtracted BLAIR sequence | Tissue properties | Dominant source(s) of contrast-BLAIR |
|---|---|---|---|---|
| 1 | DIR† | – | ρm, T1, T2 | T1, T2-BLAIR |
| lDIR | – | ρm, T1, T2 | T1, T2-BLAIR | |
| 2 | SIR† | rSIR | ρm, T1, T2, ± MT | T1, T2-BLAIR; T1, MT, T2-BLAIR |
| ENVI | – | ρm, T1, T2, ± MT | T1, T2-BLAIR; T1, MT, T2-BLAIR | |
| 3 | dSIR† | drSIR | T1, ±MT | T1, T2-BLAIR; T1, MT-BLAIR |
| cdSIR | cdrSIR | T1, ±T2, ±T2*, ±D*, ±χ, ±MT | T1, T2-BLAIR, etc. | |
| 4 | lSIR† | lrSIR | T1, ±MT | T1-BLAIR; T1, MT-BLAIR |
| clSIR | clrSIR | T1, ±T2, ±T2*, ±D*, ±χ, ±MT | T1, T2-BLAIR, etc. |
†, main sequences. BLAIR, BipoLAr Inversion Recovery; cdSIR, composite divided subtracted inversion recovery; clSIR, composite logarithmic then subtracted inversion recovery; DIR, double inversion recovery; dSIR, divided subtracted inversion recovery; ENVI, enhanced viability imaging; lDIR, logarithmic then double inversion recovery; lSIR, logarithmic then subtracted inversion recovery; MT, magnetisation transfer; SIR, subtracted inversion recovery.
Table 3
| No. | Filter, other functions | Signal equation | Figure No. |
|---|---|---|---|
| 1 | IR, TIs | STIs =1−2e−TIs/T1 | 3, 4, 5 |
| 2 | IR, TIi | STIi =1−2e−TIi/T1 | 3, 4, 5 |
| 3 | DIR, TIl, TIs | SDIR = ρm (1−2e−TIl/T1) (1−2e−TIs/T1) e−TE/T2 | |
| 4 | SIR | SSIR = STIs − STIi | |
| 5 | dSIR | 5, 6 | |
| 6 | dSIR | 5, 6 | |
| 7 | cdSIR | 9 | |
| 8 | cdSIR, SOF | SOF = ±e−∆TE/T2, ± e−∆TE/T2*, ±e−∆bD*, etc | 9 |
| 9 | dSIR, SdSIR | 9 | |
| 10 | dSIR, T1 | 9 | |
| 11 | lSIR | 7, 8 | |
| 12 | clSIR | ||
| 13 | clSIR, lSIR | , etc | |
| 14 | lSIR, dSIR† | 8 | |
| 15 | dSIR‡ | 5 | |
| 16 | lSIR§ | 8 |
†, SlSIR is the inverse hyperbolic tangent of SdSIR in the mD. ‡, is the T1 derivative of SdSIR and gives the slope of the dSIR T1-filter in the mD which is essentially constant and equal to (Figure 8). §, is the T1 derivative of the SlSIR T1-filter in the mD which varies. It has values of minus or plus infinity when and respectively (Figure 8). cdSIR, composite divided subtracted inversion recovery; clSIR, composite logarithmic then subtracted inversion recovery; DIR, double inversion recovery; dSIR, divided subtracted inversion recovery; hD, highest domain; IR, inversion recovery; lD, lowest domain; lSIR, logarithmic then subtracted inversion recovery; mD, middle domain; SIR, subtracted inversion recovery; SOF, signal from other filter; TI, inversion time.
The dSIR sequence is the BLAIR sequence most used in the present studies. T1 measurements obtained with it have been validated in quantitative phantom and human studies (16-18).
While the emphasis in this paper has been on white matter in the mD and grey matter in the hD, very high contrast can also be seen in the central grey matter of the brain when this is in the mD rather than in the hD. In addition, iron containing tissues have only been studied with dSIR using the single tissue property T1, not the composite sequence cdSIR using the two tissue properties T1 and T2*.
Understanding the contrast produced by bipolar TP-filter sequences is greatly helped by use of their TP-filters. These filters relate differences or changes in tissue properties to contrast using their slopes and explain signal and contrast patterns that can otherwise appear inexplicable using conventional approaches (19).
Artefacts from too high a value of TIi, misregistration, partial volume effects and other causes are described in reference (5). Means of avoiding these and/or reducing their effects are included in reference (5).
Tissue and fluid boundaries
dSIR and lSIR images are characterised by sharply defined high contrast, higher spatial resolution boundaries between white matter and gray matter as well as between white matter and CSF, and between gray matter and CSF. The site of the boundaries as well as their width can be changed by varying the TIs of the sequence (5). The boundaries are specific and located at iso-T1 contours. They are not part of a generalised pattern of edge enhancement.
In addition, there are well defined low signal (dark) boundaries between white matter, gray matter and CSF as modelled by bipolar TP-filters.
High signal in-plane boundaries on dSIR images usually have a uniform and consistent appearance although these are subject to partial volume effects and may simulate lesions particularly with 2D images acquired with relatively thick slices (e.g., 2–4 mm). Through-plane partial volume effects also produce high signal from boundaries and these are more prominent with thick slices.
A systematic treatment of boundaries including variation in signal with position using TP-filters, the partial derivative change in T1 with tissue fraction and the partial derivative change of tissue fraction with position is described in reference (5). Using this approach, differences in signal (i.e., contrast) can be related to differences in position within the boundary region.
Hybrid quantitative and qualitative imaging
dSIR sequences produce single images that can be used both quantitatively and qualitatively. The images show high contrast and this can be used for conventional qualitative image interpretation. The dSIR images are also T1 maps so they can be used to directly measure T1 values in regions of interest placed on them for quantitative studies. In comparison with the conventional approaches to T1 quantitation which use separate images for qualitative interpretation and for T1 mapping, dSIR images: (i) do not require working with images of two different types; (ii) allow precise placement of regions of interest (ROIs) in relation to lesions on the images used for radiological interpretation; and (iii) require no extra time for additional acquisitions.
T1 values of dSIR sequences have shown comparable accuracy to conventional T1 mapping in the mD but, as expected, differences in the lD and hD (16-18).
MT
With 2D multislice acquisitions, there may be incidental MT effects from off-resonance 180° or lower flip angle pulses used during FSE acquisitions. Using two pool modelling, MT decreases the ρm observed (ρmobs) in the free pool but also decreases the observed T1 (T1obs) of the free pool in proportion to the reduction in ρmobs (20-26). This reduction may be quite substantial and decrease normal ρmobs and T1obs by as much as 50%. In general, MT effects are decreased in disease so the reduction in ρmobs and T1obs in diseased tissue due to MT is generally less than that in normal tissue. This is manifest as an apparent increase in ρmobs and T1obs in diseased tissue relative to normal tissue. This is in addition to any increases in ρmobs and T1 in diseased tissue for other reasons such as pathological change due to oedema. To specifically recognise the MT contribution to contrast with BLAIR FSE sequences, the sequences can be designated as T1, MT-BLAIR.
2D IR FSE images with higher echo train lengths (ETLs) require more 180° or similar pulses and thus produce a corresponding increase in MT effects during multislice acquisitions. These produce a greater reduction in values of T1obs and so need shorter nulling TIs. MT effects are less with gradient echo acquisitions as often used with 3D acquisitions. As a result, T1obs may be longer with 3D IR gradient echo acquisitions than with 2D IR FSE acquisitions and so their nulling TIs may need to be correspondingly longer. The contrast between normal and abnormal tissues may also be lower with gradient echo acquisitions.
MT can be used intentionally either before or after preparation inversion pulses, as well as both before and after preparation inversion pulses in IR sequences (21). MT can also be used as an attenuating filter SOF in composite bipolar filters.
Synthetic bipolar T1-filters
Synthetic bipolar T1-filters can be of two types as shown in Table 2. The first uses Eq. [1] and the relationships shown in Figures 5,7,8 to synthesise dSIR and lSIR images with different TIs from two directly acquired images. The second type uses a synthetic bipolar TP-filter which may be the same or similar to the dSIR and lSIR filters and applies these to tissue property maps. This approach has two aspects: (i) acquiring the tissue property map which should be of high quality in the tissue property domains or parts of domains of interest; and (ii) using a visualization function such as a bipolar filter to increase contrast without saturating voxel values. These two aspects are incorporated in directly acquired dSIR and lSIR imaging.
Signs produced with dSIR sequences
The whiteout sign
This is a generalised bilateral symmetrical increase in signal in white matter of the cerebral and cerebellar hemispheres as well as the brainstem (1). The increase in signal follows the pattern of normal white matter signals and can involve all, or almost all, of the white matter in a particular slice. It has specific features such as sparing of the anterior and posterior central corpus callosum (genu and splenium) as well as peripheral white matter in the cerebral hemispheres (Figures 1,11,12,14,17,19). The white matter signal can increase up to that of the high signal boundary between white and gray matter and can be divided into five grades with grades 1 and 2 normal and grades 3, 4, 5 abnormal signs, or alternatively grade 3 as intermediate and grades 4 and 5 as abnormal white out signs (1). It is due to widespread, small increases in T1 and is thought to be particularly associated with neuroinflammation but may also be due to oedema, demyelination, degeneration and other causes. It has also been seen in mTBI, hypoxic-ischaemic injury to the brain, MS and white matter associated with tumours.
Grayout signs
With grayout signs there is a loss of contrast in gray matter such as is seen in the thalamus and basal ganglia after mTBI (Figures 11,12), or a loss of contrast between gray matter and white matter (Figure 12B). Grayout signs are generally more subtle than the whiteout sign. They are associated with an increase in T1 in gray matter.
In young subjects, the normal lateral thalamus gray matter (TGlat) tends to have a shorter T1 than the normal medial thalamus gray matter (TGmed) (upper horizontal negative solid green arrow). This leads to higher signal in the lateral thalamus with a narrow mD dSIR sequence (first vertical blue arrow) (Figure 21). The T1 of the lateral thalamus may increase in disease (middle horizontal positive green striped arrow in Figure 21) which produces a loss of contrast with the medial thalamus (blue circle). With reversion back to normal on recovery, the T1 may decrease back to normal (lowest horizontal negative green arrow) restoring contrast (second vertical blue arrow).
The whiteout sign may be accompanied by grayout signs in conditions such as mTBI as described in this paper. The term post-insult leukoencephalopathy syndromes (PILS) described in reference (1) refers to whiteout signs after mTBI, Grinker’s myelinopathy, etc. A term that includes both the whiteout sign and grayout signs as well as the possibility of ongoing effects (rather than just post-insult effects) is reactive encephalopathy (RE). Depending on the observed features (presence/absence of grayout signs), the pattern may be described as PILS/RE. Experimental studies to determine histological origins of the whiteout and grayout signs are in progress.
The bilaminar cortex sign
This is an increase in signal in the outer layer of the cortex which is often barely seen in young adults but becomes more obvious with increasing age as well as in Parkinson’s disease. It is seen with narrow mD dSIR images in which there is a high signal at the white matter gray matter junction, a lower signal layer in the inner cortex and a higher signal layer in the outer cortex (Figures 18,20). The higher signal generally corresponds to layers I–III and the lower signal to layers IV–VI of the cortex. The increase in signal is usually most obvious at the crest of gyri and to a lesser extent in the banks of sulci. It may be obscured by partial volume effects particularly on 2D images of 4 mm thickness compared with the estimated 2–4 mm thickness of the cortex. The sign may be due to increased relatively free iron content with a decrease in T1 of the normal cortex with increase in age, with an exaggeration of this in Parkinson’s disease (27). It is distinguished from the double cortex sign (28) which is due to iron deposited in a subcortical location. A bilaminar cortical appearance has been seen with MP2RAGE sequences at 7T after detailed image processing and has been attributed to differences in myelin (29). The bilaminar cortex sign differs from the high signal that can be seen at the cortical margin on T2-FLAIR images. This signal is attributed to increased T2 relative to brain in cortical veins/pia arachnoid (as opposed to a decreased T1 in the bilaminar cortex sign). It is usually more evident on 2D T2-FLAIR images with longer TIs and TRs than 3D T2-FLAIR images though the latter may have very long effective TEs. It can be present in young subjects who do not show a bilaminar cortex sign.
Bubble signs
With the narrow mD dSIR sequence nulling white matter, bubble signs are seen with an increase in T1 in white matter that crosses the high signal boundary (“overshoots”) and enters the hD (Figure 14B) (5). There is an outer high signal margin and a lower signal central region (3). Bubble signs may also be seen in the gray matter of the thalamus and basal ganglia with narrow mD dSIR sequences when the T1 in the gray matter decreases and crosses the bipolar T1-filter peak. Bubble signs of this type may become more obvious with age, and are seen in Parkinson’s disease. The decrease in T1 in gray matter sufficient to cross the high signal boundary of narrow mD dSIR sequences is illustrated in Figure 18 and probably reflects heterogeneous distribution of relatively free iron. Bubble signs may also be produced in MS by release of free iron at the rim of lesions decreasing the T1 in abnormally increased T1 lesions.
Figure 22 illustrates the origin of the bubble sign in the thalamus. The T1 of lateral gray matter is decreased from normal (TGlatnorm) to abnormal (TGlatab). In doing so, it passes through the peak signal where there is a partial volume effect between TGlatnorm and TGlatab. This results in a high signal on the periphery and a lower signal centrally in the bubble as shown by the vertical blue arrows in Figure 22.
Primary cerebral hemisphere cortices
The primary motor, somato-sensory and visual cortices layers IV–VI have relatively greater myelin than elsewhere in the cortex, and this results in a shorter T1 and so the deep cortical signal is lower than the superficial signal from layers I-III when using narrow mD dSIR sequences compared with cortex elsewhere.
Clinical use of T2-FLAIR and tissue property-BLAIR (TP-BLAIR) sequences
T2-FLAIR and T1-BLAIR sequences are complementary. The T2-FLAIR images show contrast due to larger changes in T2 and T1-BLAIR sequences such as dSIR show abnormalities due to smaller changes in T1 in white and/or gray matter in areas that appear normal or near normal with conventional T2-FLAIR sequences (e.g., Figure 14A,14B). In many diseases there are concurrent increases in both T1 and T2 so that the T1-BLAIR sequence can usefully be matched with a T2-FLAIR sequence.
Other TP-BLAIR sequences can be used besides directly acquired dSIR sequences. These include synthetic dSIRs as well as T2 and other tissue property synthetic TP-bipolar filters. Directly acquired TP-BLAIR sequences may be made sensitive to T2, T2* and D* as well as T1 as composite bipolar filters e.g., T1,T2-BLAIR.
Other sequences
The double inversion recovery (DIR) sequence
The DIR sequence multiples two IR sequences together and is one of the multiplied added subtracted and/or divided IR (MASDIR) group of sequences (5). It has a doubly negative bipolar T1-filter with two negative poles (Figure 23) (30) as well as ρm and T2-filters. Initially, the two IR sequences were used to null fat and CSF (31) but in later versions of the sequence either white or gray matter was nulled and so was CSF (32). Lesions which increase both T1 and T2 typically produce synergistic contrast. T2 preparation can be used to produce synergistic T2 contrast with that generated by the FSE data acquisition. This is synergistic with the T1 contrast of the DIR sequence (33). Incidental and/or intentional MT can also be used with the DIR sequence to increase lesion apparent ρm and T1 contrast. The log DIR (lDIR) sequence can also be used to increase lesion contrast. White matter nulled DIR sequences can be of particular value for imaging lesions in the cerebral cortex.
In a study using a gray matter and CSF nulled DIR sequence in patients with MS so that only white matter signal remained, Tillema et al. found a dark rim around lesions in white matter with longer T1s (34). The rim was due to nulling of T1s in the boundary region between normal white matter and lesions. It was evident if the T1 of the MS lesions was increased beyond the T1 of nulled gray matter. This was seen more often in MS than other disease. The nulling effect is dependent on the T1 of the mixture of tissues matching the first nulling TI of the DIR sequence. It corresponds to the high signal bubble sign seen in white matter (Figure 14B). Release of paramagnetic free iron could also have a role but the sign is due to T1 effects, not susceptibility as with the paramagnetic rim sign seen with T2*-weighted images.
SIR
Enhanced viability imaging (ENVI) is a SIR sequence which is used for increased sensitivity in the detection of Gadolinium Based Contrast Agent (GBCA) induced T1 shortening in cardiac studies (35).
MP2RAGE (36-39)
This sequence multiplies the signals of two IR sequences together and divides this by the sum of the squares of the signals from the two sequences. Complex multiplication is employed. Signal values vary between ±½. It is a negative unipolar T1-filter for brain and has a moderate negative slope between white matter and gray matter with a shallower positive slope between gray matter and CSF (Figure 24, blue curve labelled FLAWS-uni which is identical with MP2RAGE). The sequence places white matter at the upper end of the display scale, gray matter at the lower end of the display scale and CSF at an intermediate level which is similar to the divided reverse SIR (drSIR) sequence but the reverse of the dSIR sequence.
The MP2RAGE sequence can be used to display dark rims around MS lesions (41). The rims can be correlated with disease activity and are more predictive of this than paramagnetic rim signs. Similar to MP-RAGE, they reflect increases in T1 that cross the lowest signal of the MP2RAGE negative unipolar T1-filter shown in Figure 24 (blue curve). This work follows earlier studies in MS with a 3D IR gradient echo sequence (42) as well as the DIR study discussed earlier (34). The signals in these cases are from the negative poles of IR sequences and correspond with the high signal seen with dSIR sequences at the outer boundary of MS lesions using the positive pole of the dSIR sequence (Figure 14B). Two of the five high signal lesions in Figure 14B show paramagnetic rim signs in Figure 14C. In comparison with the dark rims of the MP2RAGE and 3D IR gradient echo sequence, the dSIR sequence shows a more sharply defined boundary in keeping with the shape of its T1-filter. The cut off points for showing dark rims or high signal boundaries can be changed by varying the relevant TI. There are also high signal rims seen at the margins of MS lesions which may be due to shortening of T1 in normal white matter and/or lesions by paramagnetic “free” iron [as in Fig. 6 of reference (42)]. This may be made more obvious with a drSIR sequence sensitive to decreased T1 with the longer TI nulling white matter (3).
Fluid and white matter suppression high contrast (FLAWS-hc), FLAWS-high contrast opposite (FLAWS-hco) and FLAWS-minimum (FLAWS-min) (39,40,43-45)
The FLAWS-hc and FLAWS-hco sequences have the same general mathematical form as the dSIR and drSIR sequences i.e. subtraction of two IR sequences divided by their sum i.e., , but differ in important ways: firstly, the TIs used are widely separated as opposed to dSIR and drSIR sequences where the TIs are usually narrowly separated in order to specifically target small differences or changes in T1 in tissues. Secondly, complex multiplication is used and the images are in phase-sensitive form not magnitude form. In Figure 24, the FLAWS-hc filter (green) has a positive slope over a very wide T1 Domain and the FLAWS-hco filter (orange) has a negative slope over the same very wide Domain. The magnitude of the slopes of the two filters is generally greater than that of the FLAWS-min/MP2RAGE filter. The monotonic form of the FLAWS-hc and FLAWS-hco T1-filters is strikingly different from the bipolar form of dSIR and lSIR T1-filters which have well defined, sharp discontinuities at their null points due to the use of magnitude reconstruction. They also have steep positive slopes within narrow mDs rather than the broad lower magnitude slopes of the FLAWS-hc and FLAWS-hco filters over wide T1 domains (Figure 24).
The FLAWS-min filter is the lower signal from two IR T1-filters, one nulling white matter and the other nulling CSF. This is typically divided by the sum of the signals from the two filters and has a maximum value of a ½ and a minimum value of zero for the brain. It follows the shorter TI nulling IR filter in the lower part and the longer TI nulling IR filter in the higher part of the T1 Domain. It is a positive unipolar filter for the brain with the lowest signal nulling white matter, a peak at a T1 just greater than the T1 of gray matter, and CSF zero signal (Figure 24). It has similarities to the narrow mD dSIR sequence but has a lower display range and a lower slope. Müller et al. described a halo sign when imaging MS lesions in the cortex with the FLAWS-min sequence [Fig. 5 in reference (45)]. This is due to an increase in cortex T1 in lesions passing through the peak of the FLAWS-min filter into longer T1 regions. The slope of the T1-filter of the FLAWS-min sequence around its maximum value is relatively low and this may account for the greater width of the halo sign compared with the width of the bubble sign found with the narrow mD dSIR sequence when there is an increase in T1 through the peak of that sequence (e.g., Figure 14B). There are also high signal boundaries between ventricular white matter and CSF as well as between white matter and a periventricular lesion with the FLAWS-min sequence [in Fig. 5, reference (45)] as expected from its T1-filter.
A similar pattern is seen in the leukocortical lesion shown in Figure 2B of reference (46) using T1w-FLAWS which is the same as FLAWS-min. However, the cortical lesion shown in the same figure imaged with T1w-FLAWS shows low contrast with the surrounding normal gray matter. Increased signal is seen at the boundary between gray matter and CSF. These features are consistent with Figure 24 (blue curve for FLAWS-min). FLAWS-min images do not show high signal between white matter and gray matter because the peak signal on the FLAWS-min T1-filter is not between these two tissues. If the second nulling TI (of CSF) is reduced, the FLAWS-min sequence becomes more like a narrow mD dSIR filter but with lower magnitude slopes.
In summary, the MP2RAGE and FLAWS T1-filters are unipolar or monotonic, not bipolar. They have fixed TIs and wide T1 domains with lower slopes than narrow mD dSIR bipolar T1-filters in their mDs. The FLAWS-uni/MP2RAGE T1-filter has a moderate slope, reduced CSF signal and shows white and gray matter at the upper and lower ends of the display scale. FLAWS-hc has a higher slope than FLAWS-uni/MP2RAGE, with CSF very high signal and a wide display range from WM to CSF. FLAWS-hco also has a higher magnitude negative slope than FLAWS-uni/MP2RAGE with CSF very low signal and a wide display range from CSF to white matter. FLAWS-min has a positive unipolar T1-filter with a moderate slope and peak signal at a T1 just greater than that of gray matter. The T1-filters shown in Figure 24 provide a succinct and accessible way of comparing the signal levels and contrast performance of these sequences.
White matter nulled MP-RAGE (47) and fast gray matter IR (FGATIR) (48)
These are single IR sequences that null white matter so that most of the remaining signal present in the brain is from gray matter. They have negative unipolar T1-filters. The slopes of their T1-filters are shown in Figure 5A,5B when TI is chosen to null white matter. Small differences or changes in white matter from normal show much less contrast with white matter nulled MP-RAGE than with white matter nulled narrow mD dSIR sequences (Figure 5B). Boundaries between white matter and gray matter also show higher contrast with narrow mD sequences (Figure 6).
Tissue border enhancement (TBE) (49), 3D-edge enhancing gradient echo (3D-EDGE) (50) and EDGE-MP2RAGE (51) sequences
TBE nulls signals at the junction between white and gray matter using spin echo acquisitions and was initially performed at 7T. It has a negative unipolar filter. The EDGE sequence has a single IR gradient echo and also uses a negative unipolar filter with a TI also chosen to produce zero signal at the boundary between white and gray matter with equal signals in each tissue using magnitude reconstruction. The low signal boundary seen with both techniques contains voxels with a mixture of the two tissues and a T1 between those of white and gray matter. The techniques are of value in localising boundaries and diagnosing focal cortical dysplasia. EDGE-MP2RAGE acquires two separate images, the first of which is nulled in the same way as for the 3D-EDGE sequence. FGATIR, 3D-EDGE and EDGE-MP2RAGE images can be synthetically created from MP2RAGE derived T1 maps (52).
Conclusions
BLAIR sequences provide increased lesion contrast and better defined tissue boundaries using T1, T2, MT and other tissue properties and can be readily implemented using existing sequences on MR systems. BLAIR sequences are sensitive to small changes in T1, T2 and other tissue properties and are complementary to conventional sequences such as T2-FLAIR which are sensitive to larger changes in tissue properties.
A limitation of this paper is that it is educational in content and designed to facilitate implementation of the new sequences and interpretation of the resultant images rather than to provide an account of detailed studies using the new sequences. This is because the educational aspect is seen as a necessary prelude to systematic clinical studies which will be required to establish the efficacy of BLAIR sequences.
Acknowledgments
None.
Footnote
Funding: This study was supported by
Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://qims.amegroups.com/article/view/10.21037/qims-2025-1680/coif). E.P. received fees from Bayer Swiss AG for participation on the Advisory Board, unrelated to the present work. S.J.H. received institutional technical and research support from GE HealthCare, including provision of equipment, expertise, and on-site technical support. G.M.B. is a consultant to Magnetica, Brisbane, Australia. The other authors have no conflicts of interest to declare.
Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.
Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.
References
- Condron P, Cornfeld DM, Scadeng M, Melzer TR, Newburn G, Bydder M, Kwon EE, McGeown JP, Handsfield GG, Emsden T, Tayebi M, Holdsworth SJ, Bydder GM. Ultra-High Contrast MRI: The Whiteout Sign Shown with Divided Subtracted Inversion Recovery (dSIR) Sequences in Post-Insult Leukoencephalopathy Syndromes (PILS). Tomography 2024;10:983-1013. [Crossref] [PubMed]
- Ma YJ, Shao H, Fan S, Lu X, Du J, Young IR, Bydder GM. New options for increasing the sensitivity, specificity and scope of synergistic contrast magnetic resonance imaging (scMRI) using Multiplied, Added, Subtracted and/or FiTted (MASTIR) pulse sequences. Quant Imaging Med Surg 2020;10:2030-65. [Crossref] [PubMed]
- Ma YJ, Moazamian D, Port JD, Edjlali M, Pruvo JP, Hacein-Bey L, Hoggard N, Paley MNJ, Menon DK, Bonekamp D, Pravatà E, Garwood M, Danesh-Meyer H, Condron P, Cornfeld DM, Holdsworth SJ, Du J, Bydder GM. Targeted magnetic resonance imaging (tMRI) of small changes in the T1 and spatial properties of normal or near normal appearing white and gray matter in disease of the brain using divided subtracted inversion recovery (dSIR) and divided reverse subtracted inversion recovery (drSIR) sequences. Quant Imaging Med Surg 2023;13:7304-37. [Crossref] [PubMed]
- Newburn G, McGeown JP, Kwon EE, Tayebi M, Condron P, Emsden T, Holdsworth SJ, Cornfeld DM, Bydder GM. Targeted MRI (tMRI) of Small Increases in the T1 of Normal Appearing White Matter in Mild Traumatic Brain Injury (mTBI) Using a Divided Subtracted Inversion Recovery (dSIR) Sequence. OBM Neurobiology 2023;7:201.
- Ma YJ, Moazamian D, Cornfeld DM, Condron P, Holdsworth SJ, Bydder M, Du J, Bydder GM. Improving the understanding and performance of clinical MRI using tissue property filters and the central contrast theorem, MASDIR pulse sequences and synergistic contrast MRI. Quant Imaging Med Surg 2022;12:4658-90. [Crossref] [PubMed]
- Condron P, Cornfeld DM, Bydder M, Kwon EE, Whitehead K, Pravatà E, Danesh-Meyer H, Shi C, Emsden TC, Newburn G, Scadeng M, Holdsworth SJ, Bydder GM. Ultra-High Contrast (UHC) MRI of the Brain, Spinal Cord and Optic Nerves in Multiple Sclerosis Using Directly Acquired and Synthetic Bipolar Filter (BLAIR) Images. Diagnostics (Basel) 2025;15:329. [Crossref] [PubMed]
- Newburn G, Condron P, Kwon EE, McGeown JP, Melzer TR, Bydder M, Griffin M, Scadeng M, Potter L, Holdsworth SJ, Cornfeld DM, Bydder GM. Diagnosis of Delayed Post-Hypoxic Leukoencephalopathy (Grinker's Myelinopathy) with MRI Using Divided Subtracted Inversion Recovery (dSIR) Sequences: Time for Reappraisal of the Syndrome? Diagnostics (Basel) 2024;14:418. [Crossref] [PubMed]
- Bydder M, Cornfeld DM, Melzer TR, Condron P, Newburn G, Kwon EE, Tayebi M, Scadeng M, Holdsworth SJ, Bydder GM. Log subtracted inversion recovery. Magn Reson Imaging 2025;117:110328. [Crossref] [PubMed]
- Ugurbil K. Magnetic resonance imaging at ultrahigh fields. IEEE Trans Biomed Eng 2014;61:1364-79. [Crossref] [PubMed]
- Platt T, Ladd ME, Paech D. 7 Tesla and Beyond: Advanced Methods and Clinical Applications in Magnetic Resonance Imaging. Invest Radiol 2021;56:705-25. [Crossref] [PubMed]
- Ivanov D, De Martino F, Formisano E, Fritz FJ, Goebel R, Huber L, Kashyap S, Kemper VG, Kurban D, Roebroeck A, Sengupta S, Sorger B, Tse DHY, Uludağ K, Wiggins CJ, Poser BA. Magnetic resonance imaging at 9.4 T: the Maastricht journey. MAGMA 2023;36:159-73. [Crossref] [PubMed]
- Ladd ME, Quick HH, Speck O, Bock M, Doerfler A, Forsting M, et al. Germany's journey toward 14 Tesla human magnetic resonance. MAGMA 2023;36:191-210. [Crossref] [PubMed]
- Bates S, Dumoulin SO, Folkers PJM, Formisano E, Goebel R, Haghnejad A, Helmich RC, Klomp D, van der Kolk AG, Li Y, Nederveen A, Norris DG, Petridou N, Roell S, Scheenen TWJ, Schoonheim MM, Voogt I, Webb A. A vision of 14 T MR for fundamental and clinical science. MAGMA 2023;36:211-25. [Crossref] [PubMed]
- Vachha B, Huang SY. MRI with ultrahigh field strength and high-performance gradients: challenges and opportunities for clinical neuroimaging at 7 T and beyond. Eur Radiol Exp 2021;5:35. [Crossref] [PubMed]
- Feinberg DA, Beckett AJS, Vu AT, Stockmann J, Huber L, Ma S, et al. Next-generation MRI scanner designed for ultra-high-resolution human brain imaging at 7 Tesla. Nat Methods 2023;20:2048-57. [Crossref] [PubMed]
- Bydder M, Ali F, Condron P, Cornfeld DM, Newburn G, Kwon EE, Tayebi M, Scadeng M, Melzer TR, Holdsworth SJ, Bydder GM. Validation of an ultrahigh contrast divided subtracted inversion recovery technique using a standard T1 phantom. NMR Biomed 2024;37:e5269. [Crossref] [PubMed]
- Losa L, Peruzzo D, Galbiati S, Locatelli F, Agarwal N. Enhancing T1 signal of normal-appearing white matter with divided subtracted inversion recovery: A pilot study in mild traumatic brain injury. NMR Biomed 2024;37:e5175. [Crossref] [PubMed]
- Kauppinen RA, Thotland J, Pisharady PK, Lenglet C, Garwood M. A Subtracted-Added-Divided Inversion Recovery (dSIR) Approach to Visualise the Effects of Microstructure on T1 Contrast in Human White Matter. NMR Biomed 2025;38:e70070. [Crossref] [PubMed]
- Young IR, Szeverenyi NM, Du J, Bydder GM. Pulse sequences as tissue property filters (TP-filters): a way of understanding the signal, contrast and weighting of magnetic resonance images. Quant Imaging Med Surg 2020;10:1080-120. [Crossref] [PubMed]
- Hajnal JV, Baudouin CJ, Oatridge A, Young IR, Bydder GM. Design and implementation of magnetization transfer pulse sequences for clinical use. J Comput Assist Tomogr 1992;16:7-18. [Crossref] [PubMed]
- Dortch RD, Bagnato F, Gochberg DF, Gore JC, Smith SA. Optimization of selective inversion recovery magnetization transfer imaging for macromolecular content mapping in the human brain. Magn Reson Med 2018;80:1824-35. [Crossref] [PubMed]
- Cronin MJ, Xu J, Bagnato F, Gochberg DF, Gore JC, Dortch RD. Rapid whole-brain quantitative magnetization transfer imaging using 3D selective inversion recovery sequences. Magn Reson Imaging 2020;68:66-74. [Crossref] [PubMed]
- Demir S, Clifford B, Lo WC, Tabari A, Goncalves Filho ALM, Lang M, Cauley SF, Setsompop K, Bilgic B, Lev MH, Schaefer PW, Rapalino O, Huang SY, Hilbert T, Feiweier T, Conklin J. Optimization of magnetization transfer contrast for EPI FLAIR brain imaging. Magn Reson Med 2022;87:2380-7. [Crossref] [PubMed]
- Bagnato F, Franco G, Ye F, Fan R, Commiskey P, Smith SA, Xu J, Dortch R. Selective inversion recovery quantitative magnetization transfer imaging: Toward a 3 T clinical application in multiple sclerosis. Mult Scler 2020;26:457-67. [Crossref] [PubMed]
- Kusama M, Kimura Y, Yoneyama M, Namiki T, Tamaru T, Miyagi K, Sato N. Comparison of 3D Magnetization-transfer- and Spectral-presaturation-with-inversion-recovery-based Neuromelanin Imaging. Magn Reson Med Sci 2025;24:184-90. [Crossref] [PubMed]
- Toubasi AA, Lakhani DA, Cutter G, Gheen C, Vinarsky T, Brian E, Derwenskus J, Eaton JE, Dortch RD, Xu J, Bagnato F. Improving the Detection of Myelin Integrity in Multiple Sclerosis Using Selective Inversion Recovery for MRI With Quantitative Magnetization Transfer. J Magn Reson Imaging 2025;61:2444-54. [Crossref] [PubMed]
- Nürnberger L, Gracien RM, Hok P, Hof SM, Rüb U, Steinmetz H, Hilker R, Klein JC, Deichmann R, Baudrexel S. Longitudinal changes of cortical microstructure in Parkinson's disease assessed with T1 relaxometry. Neuroimage Clin 2016;13:405-14. [Crossref] [PubMed]
- Roeben B, Zeltner L, Hagberg GE, Scheffler K, Schöls L, Bender B. Susceptibility-Weighted Imaging Reveals Subcortical Iron Deposition in PLA2G6-associated Neurodegeneration: The "Double Cortex Sign". Mov Disord 2023;38:904-6. [Crossref] [PubMed]
- Mueller SG. 7T MP2RAGE for cortical myelin segmentation: Impact of aging. PLoS One 2024;19:e0299670. [Crossref] [PubMed]
- Ma YJ, Fan S, Shao H, Du J, Szeverenyi NM, Young IR, Bydder GM. Use of Multiplied, Added, Subtracted and/or FiTted Inversion Recovery (MASTIR) pulse sequences. Quant Imaging Med Surg 2020;10:1334-69. [Crossref] [PubMed]
- Bydder GM, Young IR. MR imaging: clinical use of the inversion recovery sequence. J Comput Assist Tomogr 1985;9:659-75.
- Redpath TW, Smith FW. Technical note: use of a double inversion recovery pulse sequence to image selectively grey or white brain matter. Br J Radiol 1994;67:1258-63. [Crossref] [PubMed]
- Costagli M, Lapucci C, Zacà D, Bruschi N, Schiavi S, Castellan L, Stemmer A, Roccatagliata L, Inglese M. Improved detection of multiple sclerosis lesions with T2-prepared double inversion recovery at 3T. J Neuroimaging 2022;32:902-9. [Crossref] [PubMed]
- Tillema JM, Weigand SD, Dayan M, Shu Y, Kantarci OH, Lucchinetti CF, Port JD. Dark Rims: Novel Sequence Enhances Diagnostic Specificity in Multiple Sclerosis. AJNR Am J Neuroradiol 2018;39:1052-8. [Crossref] [PubMed]
- Foo TK, Wolff SD, Gupta SN, Kraitchman DL. Enhanced viability imaging: improved contrast in myocardial delayed enhancement using dual inversion time subtraction. Magn Reson Med 2005;53:1484-9. [Crossref] [PubMed]
- Marques JP, Kober T, Krueger G, van der Zwaag W, Van de Moortele PF, Gruetter R. MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field. Neuroimage 2010;49:1271-81. [Crossref] [PubMed]
- Kober T, Granziera C, Ribes D, Browaeys P, Schluep M, Meuli R, Frackowiak R, Gruetter R, Krueger G. MP2RAGE multiple sclerosis magnetic resonance imaging at 3 T. Invest Radiol 2012;47:346-52. [Crossref] [PubMed]
- Aichour R, Emorine T, Oubaya N, Megdiche I, Créange A, Lecler A, Kober T, Massire A, Bapst B, Improved MR. Detection of Optic Nerve Demyelination With MP2RAGE-FLAWS Compared With T 2 -Weighted Fat-Saturated Sequences. Invest Radiol 2025;60:387-96. [Crossref] [PubMed]
- Beaumont J, Saint-Jalmes H, Acosta O, Kober T, Tanner M, Ferré JC, Salvado O, Fripp J, Gambarota G. Multi T1-weighted contrast MRI with fluid and white matter suppression at 1.5 T. Magn Reson Imaging 2019;63:217-25. [Crossref] [PubMed]
- Beaumont J, Gambarota G, Saint-Jalmes H, Acosta O, Ferré JC, Raniga P, Fripp J. High-resolution multi-T1 -weighted contrast and T1 mapping with low B1>+ sensitivity using the fluid and white matter suppression (FLAWS) sequence at 7T. Magn Reson Med 2021;85:1364-78. [Crossref] [PubMed]
- Naval-Baudin P, Pons-Escoda A, Castillo-Pinar A, Martínez-Zalacaín I, Arroyo-Pereiro P, Flores-Casaperalta S, Garay-Buitron F, Calvo N, Martinez-Yélamos A, Cos M, Martínez-Yélamos S, Majós C. The T1-dark-rim: A novel imaging sign for detecting smoldering inflammation in multiple sclerosis. Eur J Radiol 2024;173:111358. [Crossref] [PubMed]
- Marshall M, Aphiwatthanasumet K, Mougin O, Stadelmann C, Morgan PS, Dineen RA, Gowland P, Evangelou N, Clarke MA. T1-Dark Rim as a Marker of New and Chronic Active Multiple Sclerosis Lesions: A Serial Study With Frequent 7T MRI. J Neuroimaging 2025;35:e70044. [Crossref] [PubMed]
- Müller J, Rahmanzadeh R, Tsagkas C, Barakovic M, Lu PJ, Yaldizli O, Weigel M, Beaumont J, La Rosa F, Bach Cuadra M, Kappos L, Kuhle J, Gambarota G, Granziera C. Cortical lesion detection using FLAWS in multiple sclerosis. Multiple Sclerosis Journal 2021;27:465.
- Beaumont J, Fripp J, Raniga P, Acosta O, Ferre JC, McMahon K, Trinder J, Kober T, Gambarota G. Multi T1-weighted contrast imaging and T1 mapping with compressed sensing FLAWS at 3 T. MAGMA 2023;36:823-36. [Crossref] [PubMed]
- Müller J, La Rosa F, Beaumont J, Tsagkas C, Rahmanzadeh R, Weigel M, Bach Cuadra M, Gambarota G, Granziera C. Fluid and White Matter Suppression: New Sensitive 3 T Magnetic Resonance Imaging Contrasts for Cortical Lesion Detection in Multiple Sclerosis. Invest Radiol 2022;57:592-600. [Crossref] [PubMed]
- Durozard P, Maarouf A, Zaaraoui W, Stellmann JP, Boutière C, Rico A, Demortière S, Guye M, Le Troter A, Dary H, Ranjeva JP, Audoin B, Pelletier J. Cortical Lesions as an Early Hallmark of Multiple Sclerosis: Visualization by 7 T MRI. Invest Radiol 2024;59:747-53. [Crossref] [PubMed]
- Saranathan M, Tourdias T, Bayram E, Ghanouni P, Rutt BK. Optimization of white-matter-nulled magnetization prepared rapid gradient echo (MP-RAGE) imaging. Magn Reson Med 2015;73:1786-94. [Crossref] [PubMed]
- Grewal SS, Middlebrooks EH, Kaufmann TJ, Stead M, Lundstrom BN, Worrell GA, Lin C, Baydin S, Van Gompel JJ. Fast gray matter acquisition T1 inversion recovery MRI to delineate the mammillothalamic tract for preoperative direct targeting of the anterior nucleus of the thalamus for deep brain stimulation in epilepsy. Neurosurg Focus 2018;45:E6. [Crossref] [PubMed]
- Costagli M, Kelley DA, Symms MR, Biagi L, Stara R, Maggioni E, Tiberi G, Barba C, Guerrini R, Cosottini M, Tosetti M. Tissue Border Enhancement by inversion recovery MRI at 7.0 Tesla. Neuroradiology 2014;56:517-23. [Crossref] [PubMed]
- Middlebrooks EH, Lin C, Westerhold E, Okromelidze L, Vibhute P, Grewal SS, Gupta V. Improved detection of focal cortical dysplasia using a novel 3D imaging sequence: Edge-Enhancing Gradient Echo (3D-EDGE) MRI. Neuroimage Clin 2020;28:102449. [Crossref] [PubMed]
- Tao S, Zhou X, Greco E, Gupta V, Freund BE, Westerhold EM, Feyissa AM, Tatum WO, Grewal S, Patel V, Middlebrooks EH. Edge-Enhancing Gradient-Echo MP2RAGE for Clinical Epilepsy Imaging at 7T. AJNR Am J Neuroradiol 2023;44:268-70. [Crossref] [PubMed]
- Middlebrooks EH, Tao S, Zhou X, Greco E, Westerhold EM, Tipton PW, Quinones-Hinojosa A, Grewal SS, Patel V. Synthetic Inversion Image Generation using MP2RAGE T1 Mapping for Surgical Targeting in Deep Brain Stimulation and Lesioning. Stereotact Funct Neurosurg 2023;101:326-31. [Crossref] [PubMed]

