Optimization of motion-corrected liver diffusion-weighted imaging at 3 Tesla (3T): incorporating complex averaging and reparametrized sinc fatsat pulse with optimized water excitation pulse
Original Article

Optimization of motion-corrected liver diffusion-weighted imaging at 3 Tesla (3T): incorporating complex averaging and reparametrized sinc fatsat pulse with optimized water excitation pulse

Zhiyong Chen1#, Zhangli Xing1#, Enshuang Zheng1, Mingcong Luo1, Caixia Fu2, Guijin Li3, Thomas Benkert4, Yunjing Xue1*, Bin Sun1*

1Department of Radiology, Fujian Medical University Union Hospital, Fuzhou, China; 2MR Application Development, Siemens Shenzhen Magnetic Resonance Ltd., Shenzhen, China; 3MR Application, Siemens Healthineers Ltd., Guangzhou, China; 4MR Application Predevelopment, Siemens Healthineers AG, Erlangen, Germany

Contributions: (I) Conception and design: Z Chen, B Sun; (II) Administrative support: Y Xue, B Sun; (III) Provision of study materials or patients: Z Chen, Z Xing, Y Xue, B Sun; (IV) Collection and assembly of data: Z Chen, Z Xing, E Zheng, M Luo, B Sun; (V) Data analysis and interpretation: Z Chen, Z Xing, C Fu, G Li, T Benkert, B Sun; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors.

#These authors contributed equally to this work.

*These authors contributed equally to this work.

Correspondence to: Yunjing Xue, MD; Bin Sun, MD. Department of Radiology, Fujian Medical University Union Hospital, 29 Xin-Quan Road, Fuzhou 350001, China. Email: xueyunjing@126.com; sunbin923@126.com.

Background: In liver diffusion-weighted imaging (DWI), single-shot echo-planar imaging (SS-EPI) sequences are susceptible to motion artifacts, resulting in image blurring and decreased lesion detection rates. This study aimed to develop and optimize a motion-corrected (MOCO) technique for liver DWI at 3 Tesla (3T). The technique incorporates motion correction, complex averaging, and a combination of a reparametrized sinc fatsat pulse with an optimized water excitation pulse.

Methods: This prospective cross-sectional study performed at Fujian Medical University Union Hospital included 42 healthy volunteers who underwent four SS-EPI DWI sequences on a 3T magnetic resonance imaging (MRI) system between January 2023 and March 2023. The sequences included a navigator-triggered (NT) MOCO-DWI, two free-breathing (FB) MOCO-DWI, and an FB conventional DWI (FB cDWI) sequence. Motion correction and complex averaging were performed for both MOCO-DWI sequences, and fat suppression was achieved using either a sinc fatsat pulse with optimized water excitation or a conventional spectral attenuated inversion recovery (SPAIR) pulse. Liver signal-to-noise ratio (SNR) was measured at b=1,000 s/mm2. Qualitative parameters were independently evaluated by three radiologists using 5-point Likert scales. Quantitative parameters were assessed using the Kolmogorov-Smirnov test, and variance homogeneity was assessed using Levene’s test. Regarding the qualitative analysis, the Friedman test was used to compare subjective scores among the four techniques.

Results: The SNRs of the liver were significantly higher with FB MOCO-DWI compared to the other EPI DWI sequences at b=1,000 s/mm2 (P<0.05). In the superior-inferior direction, the SNRs of the inferior level of the liver were higher than those of the superior level in NT MOCO-DWI. The qualitative results showed significantly higher ratings for NT MOCO-DWI and FB MOCO-DWI compared to the other EPI DWI sequences at b=1,000 s/mm2 (P<0.05). Regarding the apparent diffusion coefficient (ADC) quantification, the ADC values of the left lobe were higher than those of the right lobe in all four techniques.

Conclusions: The proposed EPI DWI technique, incorporating motion correction, complex averaging, and a modified fat suppression scheme using spectral fat saturation and binomial water excitation, was found to be clinically feasible for liver MRI. The FB MOCO-DWI sequence, with its superior SNR and excellent image quality, is recommended for liver DW imaging at 3T in clinical routine.

Keywords: Diffusion-weighted imaging (DWI); liver; free-breathing (FB); motion-corrected (MOCO); magnetic resonance imaging (MRI)


Submitted Feb 21, 2024. Accepted for publication Aug 02, 2024. Published online Aug 28, 2024.

doi: 10.21037/qims-24-340


Introduction

Diffusion-weighted imaging (DWI) has emerged as a valuable tool in clinical practice, providing a non-invasive and contrast-media free assessment of the Brownian motion of water molecules (1). The restricted motion of water in the presence of cellular structures and edema makes DWI a powerful technique for evaluating cellular density (2-4) and detecting liver lesions (5-8). For the DW sequence, the b-value represents an indicator of water diffusion capability and determines the degree of diffusion weighting. Furthermore, DWI has demonstrated utility in monitoring treatment response in liver diseases (9,10) and estimating the apparent diffusion coefficient (ADC) for qualitative and quantitative assessment (11-13).

In liver DWI, single-shot echo-planar imaging (SS-EPI) sequences are commonly employed due to their rapid acquisition and reduced susceptibility to respiratory and cardiac motion artifacts (14). Free-breathing (FB) without respiratory triggering is commonly used in clinical practice due to its high signal-to-noise ratio (SNR)-time efficiency. However, the motion occurring between different b-values or averages can compromise the quality of the resulting DWI images and ADC maps. This can lead to image blurring, less reliable ADC maps, and consequently reduced lesion detection rates. Various respiratory compensation techniques, such as breath-holding (BH) and respiratory/navigator-triggered (NT) acquisition, have been investigated to improve ADC measurement reproducibility, SNR, and image quality in liver DWI (15-18). However, the implementation of BH decreases patient comfort, whereas respiratory/NT prolongs the measurement time. To address this limitation, a novel SS-EPI DWI sequence with motion correction (19) was developed for abdominal application. Additionally, complex averaging (20), and a combination of a reparametrized sinc fatsat pulse with an optimized water excitation pulse was proposed for further improving the image qualities of DWI (21,22).

Hence, the objective of this study was to develop an FB-optimized liver DWI protocol incorporating motion correction, complex averaging, and a novel fat suppression scheme, and to assess its image quality in comparison to various DWI protocols. We present this article in accordance with the STROBE reporting checklist (available at https://qims.amegroups.com/article/view/10.21037/qims-24-340/rc).


Methods

Study participants

This prospective, cross-sectional study was conducted in accordance with the Declaration of Helsinki (as revised in 2013) and approved by the Fujian Medical University Union Hospital (No. 2023-KY018), and written informed consent was provided by all participants. A total of 42 healthy volunteers [22 males and 20 females, mean age, 28.7±4.2 years (range, 22–39 years)] were enrolled for magnetic resonance (MR) imaging of the liver at Fujian Medical University Union Hospital between January 2023 and March 2023. The inclusion criteria for this study were as follows: adults with no history of liver disease. The exclusion criteria were as follows: (I) imaging evidence of liver disease, including focal or diffuse liver lesions detected by liver MR imaging (MRI); (II) contraindications to MRI; (III) physical complaints during examination. The enrollment process of healthy volunteers is shown in Figure 1.

Figure 1 Flow diagram shows the enrollment of healthy volunteers. MRI, magnetic resonance imaging; NT, navigator trigger; FB, free breathing; MOCO, motion-corrected; DWI, diffusion-weighted imaging; SPAIR, spectral attenuated inversion recovery; cDWI, conventional diffusion-weighted imaging.

MRI protocol

MRI examinations were conducted using a 3 Tesla (3T) MR scanner (MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany), with the utilization of an 18-channel body matrix coil combined with a 32-channel spine matrix coil. All four DWI sequences, which encompassed three newly developed motion-corrected (MOCO)-DWI sequences and an FB conventional DWI (FB cDWI), were performed in a random order. The detailed scan parameters are presented in Table 1. For the three novel MOCO-DWI sequences, a non-rigid in-plane registration technique, named the inverse consistent deformable registration algorithm (19), was employed between averages to mitigate motion-related blurring. Following complex-valued averaging to reduce Rician noise bias (17), the same registration algorithm was performed for different b-values. Fat suppression was achieved by combining a binomial water excitation pulse (1-3-3-1 scheme, monopolar encoding) with a reparametrized sinc-shaped spectral fatsat pulse, or alternatively, a conventional spectral attenuated inversion recovery (SPAIR) pulse. In the case of cDWI, fat suppression was implemented using a conventional SPAIR pulse, and neither motion correction nor standard magnitude-based averaging was applied.

Table 1

Scan parameters for cDWI and MOCO-DWI

Parameters FB MOCO-DWI NT MOCO-DWI FB MOCO-DWI SPAIR FB cDWI
Breathing scheme FB NT FB FB
Field of view (mm) 380×380 380×380 380×380 380×380
Matrix 160×160 160×160 160×160 160×160
FOV phase 81.30% 81.30% 81.30% 81.30%
TR (msec) 4,700 1,100 4,700 4,700
TE (msec) 54 54 54 54
Bandwidth (Hz/pixel) 2,404 2,404 2,404 2,404
EPI factor 130 130 130 130
In-plane parallel imaging GRAPPA 2 GRAPPA 2 GRAPPA 2 GRAPPA 2
b-value (s/mm2) 0/500/1,000 0/1,000 0/500/1,000 0/500/1,000
Averages for b-values 3/5/18 3/9 3/5/18 3/5/18
Voxel (mm3) 1.2×1.2×5.0 1.2×1.2×5.0 1.2×1.2×5.0 1.2×1.2×5.0
Slice number 27 27 27 27
Fat saturation technique FatSat + water excitation FatSat + water excitation SPAIR SPAIR
Acquisition time (sec) 138 136.15±14.23 138 138

cDWI, conventional diffusion-weighted imaging; MOCO, motion-corrected; DWI, diffusion-weighted imaging; FB, free breathing; NT, navigator trigger; SPAIR, spectral attenuated inversion recovery; FOV, field of view; TR, repetition time; TE, echo time; EPI, echo-planar imaging; GRAPPA, generalized autocalibrating partial parallel acquisition.

Image analysis

Qualitative image analysis

Three board-certified abdominal radiologists, with 15, 20, and 30 years of experience in abdominal MRI, respectively, performed an independent evaluation of all echo-planar imaging (EPI) DWI images. The images were anonymized and presented to the reviewers in a random order, without any information regarding the acquisition methods or imaging parameters. For follow-up analysis, the radiologists assessed the overall image quality and sharpness of liver contours at b-values of 0 and 1,000 s/mm2 using a 5-point Likert scale, transforming the qualitative evaluation into a specific quantitative value: 1= nondiagnostic quality; 2= substantial deficits in image quality; 3= moderate image quality; 4= good image quality; and 5= excellent image quality. They also rated the severity of artifacts at b-values of 0 and 1,000 s/mm2 as follows: 1= not diagnostic; 2= severe artifacts; 3= moderate artifacts; 4= mild artifacts; and 5= no artifacts. Furthermore, the effect of the heartbeat in the left liver lobe only at b-values of 1,000 s/mm2 was rated as 1= nondiagnostic quality; 2= substantial impact on image quality; 3= moderate impact on image quality; 4= little impact on image quality; and 5= no artifacts.

Quantitative image analysis

A reader with eight years of experience conducted quantitative analysis on the source images and assessed the liver SNR at a b-value of 1,000 s/mm2. The reader, who was blinded to the acquisition scheme, placed round regions of interest (ROIs) with an area of 1.5–2.5 cm2 in the right and left liver lobes, carefully avoiding blood vessels, artifacts, and lesions. Three representative sections were selected for each DWI sequence in the superior-inferior direction. The central section corresponded to the level of the right portal vein, whereas the superior and inferior sections were obtained three or four section levels above or below the central section.

The reader measured the mean signal intensity (SI) of the liver ROIs and the SI of an equally sized ROI placed in the background (air) region of the corresponding section, close to the liver ROI site, ensuring the absence of prominent artifacts. This procedure was repeated, with similar background ROIs placed in the exact position across all other image slices. For the calculation of SNR, the following equation was utilized (23):

SNR=SI/(RfSDbg)

Where SI represents the SI of a specific ROI, standard deviation of background (SDbg) corresponds to the standard deviation of the selected artifact-free ROI placed on the background (air) of the slice, and Rf denotes the Rayleigh correction factor (0.655) used to convert the Rayleigh distribution of MRI signals in the vicinity of 0 (background) to a Gaussian distribution. In each volunteer, the location and size of the ROI remained consistent across all EPI DWI sequences (b=1,000 s/mm2) and their corresponding ADC maps.

Statistical analysis

The statistical analysis for this study involved both qualitative and quantitative parameters. The Friedman test, a nonparametric test, was used to compare qualitative parameters among the four sequences. For quantitative parameters, such as ADC and SNRs, normality was assessed using the Kolmogorov-Smirnov test, and variance homogeneity was assessed using Levene’s test.

For the quantitative analysis, the independent samples t-test was used to compare ADC differences between the left and right lobes of the liver. One-way analysis of variance (ANOVA) test was performed to compare ADC differences among the superior, middle, and inferior levels of the liver within each technique. The SNRs of the four techniques were compared using the ANOVA test, followed by Dunn-Bonferroni post-hoc test for significant pairwise comparisons. SNRs of the left and right lobes of the liver, as well as the superior, middle, and inferior levels of the liver within each technique, were also compared using independent samples t-test and ANOVA test, separately.

Regarding the qualitative analysis, the Friedman test was used to compare subjective scores among the four techniques. The Dunn-Bonferroni post-hoc test was applied to adjust for all significant pairwise comparisons. Inter-reader agreement for scores was measured using Fleiss kappa statistics, with values ranging from poor to excellent indicating the level of agreement between readers.

The statistical analysis was conducted using the software SPSS 26.0 (IBM Corp., Armonk, NY, USA). Differences were considered statistically significant when the two-sided P values were less than 0.05.


Results

All DWI sequences were successfully acquired in all cases. The mean scan time for the NT MOCO-DWI (136.15±14.23 s) sequence was not significantly different compared to the FB MOCO-DWI, FB MOCO-DWI SPAIR, and FB cDWI sequences (138 s) (P=0.538).

Regarding the comparison of ADC values among the four techniques, for the left–right direction, the ADC values of the left lobe were higher than those of the right lobe in all four techniques (FB MOCO-DWI: P<0.001, NT MOCO-DWI: P=0.003, FB MOCO-DWI SPAIR: P=0.007, FB cDWI: P=0.012). However, for the superior-inferior direction, all four techniques showed no significant difference (FB MOCO-DWI: P=0.455, NT MOCO-DWI: P=0.292, FB MOCO-DWI SPAIR: P=0.457, FB cDWI: P=0.670) (Table 2).

Table 2

Comparison of ADC values of different anatomical regions with four techniques

Technique Left lobe Right lobe P value Superior level Middle level Inferior level P value
FB MOCO-DWI 1,138.32±111.12 1,066.04±70.76 <0.001 1,129.56±110.43 1,078.72±91.38 1,076.58±82.76 0.455
NT MOCO-DWI 1,119.52±114.1,1 1,055.78±63.01 0.003 1,104.06±78.38 1,075.63±76.73 1,064.15±95.75 0.292
FB MOCO-DWI SPAIR 1,207.71±254.55 1,092.65±74.41 0.007 1,134.20±97.29 1,121.99±89.82 1,159.83±93.06 0.457
FB cDWI 1,190.75±157.83 1,099.69±163.86 0.012 1,157.46±176.55 1,144.36±177.19 1,106.53±165.63 0.670

Data are presented as mean ± standard deviation. ADC values were given in ×10−6 mm2/s. Mean ADC values measured in different anatomical regions of liver. ADC, apparent diffusion coefficient; FB, free breathing; MOCO, motion-corrected; DWI, diffusion-weighted imaging; NT, navigator trigger; SPAIR, spectral attenuated inversion recovery; cDWI, conventional diffusion-weighted imaging.

In terms of SNRs, FB MOCO-DWI demonstrated significantly higher SNRs of the liver compared to the other techniques (P<0.001), including the whole liver, left lobe, right lobe, and superior, middle, and inferior levels of the liver (Table 3, Figure 2). There were no significant differences in SNRs between the left and right lobes for all four techniques. In the superior-inferior direction, the SNRs of the inferior level of the liver were higher than those of the superior level in NT MOCO-DWI (P=0.041) (Table 4).

Table 3

Comparison of SNR with four techniques

Parameters FB MOCO-DWI NT MOCO-DWI FB MOCO-DWI SPAIR FB cDWI P value
Whole liver 93.57±46.61 65.65±26.31 65.18±25.74 53.52±16.86 <0.001
Left lobe 95.52±50.71 63.76±26.24 65.65±29.38 53.10±17.46 <0.001
Right lobe 91.61±45.42 67.549±28.47 64.72±25.54 54.96±17.49 <0.001
Superior level 82.17±40.81 58.08±24.13 60.39±23.78 50.72±17.29 <0.001
Middle level 96.80±55.48 63.74±26.24 66.03±29.22 54.84±18.19 <0.001
Inferior level 101.73±52.47 71.29±31.78 69.13±28.91 55.01±17.70 <0.001

Data are presented as mean ± standard deviation. P1, FB cDWI vs. SPAIR; P2, FB cDWI vs. NT MOCO-DWI; P3, FB cDWI vs. FB MOCO-DWI; P4, SPAIR vs. NT MOCO-DWI; P5, SPAIR vs. FB MOCO-DWI; P6, NT MOCO-DWI vs. FB MOCO-DWI. Whole liver: P1=0.001, P2<0.001, P3<0.001, P4=0.780, P5<0.001, P6<0.001. Left lobe: P1=0.004, P2=0.005, P3<0.001, P4=0.944, P5<0.001, P6<0.001. Right lobe: P1=0.003, P2=0.003, P3<0.018, P4=0.442, P5<0.001, P6<0.001. Superior level: P1=0.002, P2=0.108, P3<0.001, P4=0.162, P5=0.012, P6=0.001. Middle level: P1=0.008, P2=0.006, P3<0.001, P4=0.485, P5=0.003, P6<0.001. Inferior level: P1=0.002, P2=0.001, P3<0.001, P4=0.889, P5<0.001, P6<0.001. SNR, signal-to-noise ratio; FB, free breathing; MOCO, motion-corrected; DWI, diffusion-weighted imaging; NT, navigator trigger; SPAIR, spectral attenuated inversion recovery; cDWI, conventional diffusion-weighted imaging.

Figure 2 A violin plot of SNR in whole liver measured with different DWI sequences. SNR, signal-to-noise ratio; FB, free breathing; MOCO, motion-corrected; DWI, diffusion-weighted imaging; NT, navigator trigger; SPAIR, spectral attenuated inversion recovery; cDWI, conventional diffusion-weighted imaging.

Table 4

Comparison of SNR values of different anatomical regions with four techniques

Technique Left lobe Right lobe P value Superior level Middle level Inferior level P value Pa Pb Pc
FB MOCO-DWI 95.52±50.71 91.61±45.42 0.304 82.17±40.81 96.80±55.48 101.73±52.47 0.188 0.656 0.079 0.188
NT MOCO-DWI 63.76±26.24 67.54±28.47 0.118 58.08±24.13 60.39±23.78 71.29±31.78 0.108 0.140 0.041 0.563
FB MOCO-DWI SPAIR 65.65±29.38 64.72±25.55 0.763 60.39±23.78 66.03±29.22 69.13±28.91 0.346 0.354 0.152 0.610
FB cDWI 53.10±17.46 54.96±17.49 0.547 50.72±17.29 54.84±18.19 55.01±17.70 0.467 0.296 0.277 0.966

Data are presented as mean ± standard deviation. Pa, superior level vs. middle level; Pb, superior level vs. inferior level; Pc, middle level vs. inferior level. SNR, signal-to-noise ratio; FB, free breathing; MOCO, motion-corrected; DWI, diffusion-weighted imaging; NT, navigator trigger; SPAIR, spectral attenuated inversion recovery; cDWI, conventional diffusion-weighted imaging.

Regarding image quality assessment, there was excellent agreement among the three readers (Fleiss Kappa value: 0.674–0.941). All readers consistently rated FB MOCO-DWI as having better overall image quality, sharper liver contour, milder artifacts, and less cardiac impact compared to the other techniques (P<0.001) (Table 5, Figure 3).

Table 5

Results of qualitative analysis

Criteria FB MOCO-DWI NT MOCO-DWI FB MOCO-DWI SPAIR FB cDWI P(b=0) P(b=1,000)
b=0 b=1,000 b=0 b=1,000 b=0 b=1,000 b=0 b=1,000
Overall image quality
   R1 4.91±0.30 4.71±0.46 4.90±0.35 4.43±0.55 4.71±0.46 3.83±0.58 4.36±0.69 3.57±0.67 <0.001 <0.001
   R2 4.93±0.26 4.70±0.47 4.83±0.38 4.43±0.55 4.79±0.42 3.86±0.57 4.38±0.70 3.57±0.63 <0.001 <0.001
   R3 4.95±0.22 4.70±0.47 4.83±0.36 4.43±0.54 4.79±0.42 3.86±0.57 4.33±0.69 3.54±0.50 <0.001 <0.001
   Average 4.92±0.27 4.70±0.46 4.84±0.37 4.43±0.54 4.76±0.43 3.85±0.57 4.35±0.69 3.56±0.60
   Kappa 0.783 0.812 0.941 0.878 0.869 0.872 0.832 0.765
Sharpness of liver contour
   R1 4.83±0.38 4.59±0.59 4.95±0.22 4.76±0.53 4.31±0.47 3.62±0.73 3.98±0.56 3.29±0.60 <0.001 <0.001
   R2 4.88±0.33 4.52±0.59 4.95±0.22 4.67±0.57 4.38±0.49 3.60±0.70 4.02±0.60 3.31±0.60 <0.001 <0.001
   R3 4.88±0.33 4.52±0.59 4.93±0.26 4.67±0.57 4.40±0.50 3.62±0.70 4.02±0.56 3.36±0.58 <0.001 <0.001
   Average 4.86±0.34 4.54±0.59 4.94±0.23 4.70±0.55 4.37±0.48 3.61±0.70 4.01±0.57 3.31±0.59
   Kappa 0.796 0.908 0.849 0.761 0.863 0.921 0.885 0.854
Severity of artifacts
   R1 4.98±0.15 4.95±0.22 4.75±0.22 4.89±0.33 4.33±0.48 3.62±0.58 4.12±0.50 3.71±0.51 <0.001 <0.001
   R2 4.95±0.22 4.90±0.30 4.93±0.26 4.79±0.42 4.33±0.48 3.60±0.54 4.17±0.54 3.71±0.51 <0.001 <0.001
   R3 4.98±0.15 4.93±0.26 4.93±0.26 4.79±0.42 4.36±0.48 3.62±0.58 4.17±0.53 3.64±0.49 <0.001 <0.001
   Average 4.97±0.18 4.92±0.13 4.93±0.24 4.82±0.39 4.34±0.48 3.61±0.57 4.15±0.52 3.69±0.50
   Kappa 0.742 0.674 0.733 0.681 0.894 0.851 0.680 0.758
Effects of the heartbeat
   R1 4.60±0.66 4.31±0.78 3.64±0.76 3.45±0.77 <0.001
   R2 4.64±0.58 4.31±0.78 3.64±0.75 3.41±0.76 <0.001
   R3 4.62±0.58 4.36±0.79 3.67±0.79 3.45±0.77 <0.001
   Average 4.62±0.60 4.32±0.78 3.65±0.76 3.43±0.76
   Kappa 0.796 0.844 0.872 0.817

Mean ± standard for the mean image quality ratings calculated are given separately for the two different b-values maps. 5= excellent, 4= good, 3= moderate, 2= fair, 1= nondiagnostic. On the right, P values for the overall comparisons using the Friedman test are given when b=0 and b1,000. If the Friedman test showed statistical significance, the Dunn-Bonferroni post-hoc test for all pairwise comparisons was performed. P1–P12 values were presented only when they were less than 0.05. P1, b=0, FB cDWI vs. SPAIR; P2, b=50, FB cDWI vs. NT MOCO-DWI; P3, b=0, FB cDWI vs. FB MOCO-DWI; P4, b=0, SPAIR vs. NT MOCO-DWI; P5, b=0, SPAIR vs. FB MOCO-DWI; P6, b=0, NT MOCO-DWI vs. FB MOCO-DWI; P7, b=1,000, FB cDWI vs. SPAIR; P8, b=1,000, FB cDWI vs. NT MOCO-DWI; P9, b=1,000, FB cDWI vs. FB MOCO-DWI; P10, b=1,000, SPAIR vs. NT MOCO-DWI; P11, b=1,000, SPAIR vs. FB MOCO-DWI; P12, b=1,000, NT MOCO-DWI vs. FB MOCO-DWI. Overall image quality: for Reader 1: P1=0.027, P2=0.002, P3<0.001, P4=0.370, P5=0.168, P6=0.629, P7=0.202, P8<0.001, P9<0.001, P10=0.001, P11<0.001, P12=0.098. For Reader 2: P1=0.011, P2=0.006, P3<0.001, P4=0.836, P5=0.301, P6=0.408, P7=0.202, P8<0.001, P9<0.001, P10=0.001, P11<0.001, P12=0.157. For Reader 3: P1=0.004, P2=0.002, P3<0.001, P4=0.207, P5=0.168, P6=0.352, P7=0.285, P8<0.001, P9<0.001, P10=0.001, P11<0.001, P12=0.121. Sharpness of liver contour: for Reader 1: P1=0.067, P2<0.001, P3<0.001, P4<0.001, P5<0.001, P6=0.490, P7=0.067, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.490. For Reader 2: P1=0.058, P2<0.001, P3<0.001, P4<0.001, P5<0.001, P6=0.679, P7=0.112, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.581. For Reader 3: P1=0.025, P2<0.001, P3<0.001, P4<0.001, P5=0.001, P6=0.809, P7=0.168, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.605. Severity of artifacts: for Reader 1: P1=0.202, P2<0.001, P3<0.001, P4<0.001, P5<0.001, P6=0.863, P7=0.704, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.605. For Reader 2: P1=0.352, P2<0.001, P3<0.001, P4<0.001, P5<0.001, P6=0.863, P7=0.535, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.427. For Reader 3: P1=0.202, P2<0.001, P3<0.001, P4<0.001, P5<0.001, P6=0.730, P7=0.890, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.388. Effects of the heartbeat: for Reader 1: P7=0.270, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.157. For Reader 2: P7=0.190, P8<0.001, P9<0.001, P10=0.001, P11<0.001, P12=0.084. For Reader 3: P7=0.301, P8<0.001, P9<0.001, P10<0.001, P11<0.001, P12=0.168. FB, free breathing; MOCO, motion-corrected; DWI, diffusion-weighted imaging; NT, navigator trigger; SPAIR, spectral attenuated inversion recovery; cDWI, conventional diffusion-weighted imaging.

Figure 3 Comparisons of subjective image analysis and SNRs of the four EPI DWI (A) with the corresponding ADC maps (B) at b=1,000 s/mm2. FB MOCO-DWI demonstrated significantly higher SNRs of the liver compared to the other techniques (P<0.001), including the whole liver, left lobe, right lobe, and superior, middle, and inferior levels of the liver, and demonstrated significantly better the overall image quality, the liver contour’s sharpness, the artifacts’ severity, and the heartbeat’s effects compared to the other DWI sequences, except for NT MOCO-DWI. FB, free breathing; MOCO, motion-corrected; DWI, diffusion-weighted imaging; NT, navigator trigger; SPAIR, spectral attenuated inversion recovery; cDWI, conventional diffusion-weighted imaging; SNR, signal-to-noise ratio; EPI, echo-planar imaging; ADC, apparent diffusion coefficient.

Discussion

In this study, the researchers aimed to evaluate a novel MOCO liver diffusion imaging method to address motion artifacts and improve image quality. They compared four EPI DWI protocols in terms of acquisition time, SNR, and image quality. The results showed that the SNR of the liver varied significantly among the four sequences, with the FB MOCO-DWI using optimized water excitation pulse performing the best.

The use of motion correction and complex averaging in MOCO-DWI contributed to an increase in SNR compared to FB cDWI. Previous studies have also reported the benefits of MOCO-DWI in improving SNR, such as in diffusion-weighted MRI of prostate cancer (20). The authors additionally attributed the higher SNR in FB MOCO-DWI to the application of fat suppression using a binomial water excitation pulse combined with a sinc-shaped spectral fatsat pulse, which did not cause signal reduction compared to using a SPAIR pulse. The SPAIR pulse’s spectral width can lead to partial water saturation and decreased SNR.

Comparing FB MOCO-DWI with NT MOCO-DWI, the former showed a significantly higher SNR in the liver. This can be attributed to the larger average number used in FB MOCO-DWI, which not only improves the SNR but also increases the likelihood of acquiring data during the diastole phase, reducing signal loss related to cardiac motion, especially in the left lobe of the liver (24). Despite similar acquisition times for the two MOCO-DWI sequences, FB MOCO-DWI was more time-efficient than NT MOCO-DWI because it acquired three b-values compared to two b-values in NT MOCO-DWI. However, it is important to note that the additional b500 image in FB MOCO-DWI does not impact the ADC calculation, as it lies between the b0 and b1,000 values (25). Overall, the study demonstrated the advantages of the FB MOCO-DWI sequence with optimized water excitation pulse in terms of SNR and image quality, making it a promising approach for liver diffusion imaging.

In liver imaging with DWI, respiratory and cardiac motion-induced artifacts can still be problematic, especially at 3T. To address these challenges, respiratory triggers (RTs) (18,26,27) and NTs (18,28) have been utilized. Previous studies have shown that RT and NT DWI can improve image sharpness compared to untriggered FB DWI in healthy volunteers (29). Several published reports have suggested that RT or NT acquisition can enhance liver lesion detection compared to FB acquisition, which is more susceptible to respiratory misregistration (26,30). However, a study by Choi et al. (16) found that FB and NT DWI had equal performance in assessing lesion detection and characterization for 3.0T hepatic MRI, with FB acquisition being more time-efficient than NT acquisition. A liver intravoxel incoherent motion (IVIM) imaging study by Li et al. (31) based on 22 healthy participants reported that FB can be used to replace RT due to slightly better reproducibility, while saving approximately half of the image data acquisition time. Wang et al. (32) demonstrated that an FB technique for simultaneous water-fat separation and T1 mapping of the whole liver (SWALI) in a single scan can achieve 2.5 mm isotropic spatial resolution, superior to conventional liver T1 mapping techniques under BH conditions.

In the present study, the results showed that FB MOCO-DWI demonstrated significantly better in terms of the overall image quality, the liver contour sharpness, the artifact severity, and the heartbeat effects compared to the other DWI sequences except NT MOCO-DWI. The motion-correction employed in FB MOCO-DWI contributed to reducing motion-induced artifacts. Additionally, the use of complex averaging technique and the new fat suppression scheme provided higher SNR and more homogeneity for liver signal and improved visualization of liver contours. Although there was a statistically significantly higher SNR at b=1,000 s/mm2 for FB MOCO-DWI compared to NT MOCO-DWI, this difference was not statistically significant in qualitative analysis. This may be attributed to the fact that NT acquisition mitigates respiratory motion-induced artifacts, resulting in less pronounced differences in appearance between the two techniques. Overall, the study highlights the benefits of FB MOCO-DWI in terms of improved image quality and reduced motion artifacts.

In this study, the mean ADC values were found to decrease from the left to the right hepatic lobe, which was consistent with previous studies (33,34). This observation suggests that cardiac motion can influence the ADC values, particularly in the left hepatic lobe (35,36). However, no statistically significant differences in mean ADC values were observed in the cranial to caudal directions among the four DWI sequences.

It is important to acknowledge the limitations of the study. Firstly, the diagnostic performance of the new MOCO-DWI technique was not evaluated in clinical practice, so its effectiveness in detecting and characterizing liver lesions remains to be determined. Secondly, the study did not assess the reproducibility of ADC values in each of the four sequences, which would provide insights into the reliability and consistency of the measurements. Lastly, the new MOCO-DWI sequence using SS-EPI is still susceptible to susceptibility artifacts, resulting in geometric distortion, signal attenuation, and image blurring (37). These artifacts can potentially affect the accuracy and interpretation of the images.

In conclusion, the proposed EPI DWI technique, incorporating motion correction, complex averaging, and a modified fat suppression scheme, demonstrated clinical feasibility for liver MRI. Among the four sequences evaluated, FB MOCO-DWI showed superior SNR and excellent image quality, making it a recommended choice for liver DW imaging at 3T in clinical routine.


Acknowledgments

Funding: None.


Footnote

Reporting Checklist: The authors have completed the STROBE reporting checklist. Available at https://qims.amegroups.com/article/view/10.21037/qims-24-340/rc

Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://qims.amegroups.com/article/view/10.21037/qims-24-340/coif). C.F. is an employee of Siemens Shenzhen Magnetic Resonance Ltd. G.L. is an employee of Siemens Healthineers Ltd. T.B. is an employee of Siemens Healthineers AG. The other authors have no conflicts of interest to declare.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013). The study was approved by Fujian Medical University Union Hospital (No. 2023-KY018) and informed consent was provided by all individual participants.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Koh DM, Collins DJ. Diffusion-weighted MRI in the body: applications and challenges in oncology. AJR Am J Roentgenol 2007;188:1622-35. [Crossref] [PubMed]
  2. Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis E, Laval-Jeantet M. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology 1986;161:401-7. [Crossref] [PubMed]
  3. Bammer R. Basic principles of diffusion-weighted imaging. Eur J Radiol 2003;45:169-84. [Crossref] [PubMed]
  4. Le Bihan D, Johansen-Berg H. Diffusion MRI at 25: exploring brain tissue structure and function. Neuroimage 2012;61:324-41. [Crossref] [PubMed]
  5. Galea N, Cantisani V, Taouli B. Liver lesion detection and characterization: role of diffusion-weighted imaging. J Magn Reson Imaging 2013;37:1260-76. [Crossref] [PubMed]
  6. Zhang Y, Chen J, Yang C, Dai Y, Zeng M. Preoperative prediction of microvascular invasion in hepatocellular carcinoma using diffusion-weighted imaging-based habitat imaging. Eur Radiol 2024;34:3215-25. [Crossref] [PubMed]
  7. Taron J, Johannink J, Bitzer M, Nikolaou K, Notohamiprodjo M, Hoffmann R. Added value of diffusion-weighted imaging in hepatic tumors and its impact on patient management. Cancer Imaging 2018;18:10. [Crossref] [PubMed]
  8. Huang J, Leporq B, Hervieu V, Dumortier J, Beuf O, Ratiney H. Diffusion-Weighted MRI of the Liver in Patients With Chronic Liver Disease: A Comparative Study Between Different Fitting Approaches and Diffusion Models. J Magn Reson Imaging 2024;59:894-906. [Crossref] [PubMed]
  9. Bharwani N, Koh DM. Diffusion-weighted imaging of the liver: an update. Cancer Imaging 2013;13:171-85. [Crossref] [PubMed]
  10. Wagner M, Doblas S, Daire JL, Paradis V, Haddad N, Leitão H, Garteiser P, Vilgrain V, Sinkus R, Van Beers BE. Diffusion-weighted MR imaging for the regional characterization of liver tumors. Radiology 2012;264:464-72. [Crossref] [PubMed]
  11. Wáng YXJ, Zhao KX, Ma FZ, Xiao BH. The contribution of T2 relaxation time to MRI-derived apparent diffusion coefficient (ADC) quantification and its potential clinical implications. Quant Imaging Med Surg 2023;13:7410-6. [Crossref] [PubMed]
  12. Surov A, Eger KI, Potratz J, Gottschling S, Wienke A, Jechorek D. Apparent diffusion coefficient correlates with different histopathological features in several intrahepatic tumors. Eur Radiol 2023;33:5955-64. [Crossref] [PubMed]
  13. Tang Z, Wu YP, Tan BG, Chen XQ, Guo WW, Wu KS, Zhang XM, Chen TW, Zhou HY. Apparent diffusion coefficient and its standard deviation from diffusion-weighted imaging in preoperative predicting liver invasion by T3-staged resectable gallbladder carcinoma. Clin Radiol 2024;79:e247-55. [Crossref] [PubMed]
  14. Lewis S, Dyvorne H, Cui Y, Taouli B. Diffusion-weighted imaging of the liver: techniques and applications. Magn Reson Imaging Clin N Am 2014;22:373-95. [Crossref] [PubMed]
  15. Chen X, Qin L, Pan D, Huang Y, Yan L, Wang G, Liu Y, Liang C, Liu Z. Liver diffusion-weighted MR imaging: reproducibility comparison of ADC measurements obtained with multiple breath-hold, free-breathing, respiratory-triggered, and navigator-triggered techniques. Radiology 2014;271:113-25. [Crossref] [PubMed]
  16. Choi JS, Kim MJ, Chung YE, Kim KA, Choi JY, Lim JS, Park MS, Kim KW. Comparison of breathhold, navigator-triggered, and free-breathing diffusion-weighted MRI for focal hepatic lesions. J Magn Reson Imaging 2013;38:109-18. [Crossref] [PubMed]
  17. Larsen NE, Haack S, Larsen LP, Pedersen EM. Quantitative liver ADC measurements using diffusion-weighted MRI at 3 Tesla: evaluation of reproducibility and perfusion dependence using different techniques for respiratory compensation. MAGMA 2013;26:431-42. [Crossref] [PubMed]
  18. Taouli B, Sandberg A, Stemmer A, Parikh T, Wong S, Xu J, Lee VS. Diffusion-weighted imaging of the liver: comparison of navigator triggered and breathhold acquisitions. J Magn Reson Imaging 2009;30:561-8. [Crossref] [PubMed]
  19. Jolly M P, Guetter C, Guehring J. Cardiac Segmentation in MR Cine Data Using Inverse Consistent Deformable Registration. 2010 IEEE International Symposium on Biomedical Imaging: From Nano to Macro, Rotterdam, Netherlands, 2010:484-7.
  20. Kordbacheh H, Seethamraju RT, Weiland E, Kiefer B, Nickel MD, Chulroek T, Cecconi M, Baliyan V, Harisinghani MG. Image quality and diagnostic accuracy of complex-averaged high b value images in diffusion-weighted MRI of prostate cancer. Abdom Radiol (NY) 2019;44:2244-53. [Crossref] [PubMed]
  21. Chebrolu VV, Frick M, Kollasch P, Schmitt B, Deshpande V, Grinstead J, Spence D, Felmlee J, Amrami KK. Fat Saturation in Musculoskeletal MRI using Spectral Heterogeneity Adaptive RF Pulses (SHARP). Proceedings Joint Annual Meeting ISMRM-ESMRMB 2018.
  22. Sauer ST, Christner SA, Schlaiß T, Metz C, Schmid A, Kunz AS, Pabst T, Weiland E, Benkert T, Bley TA, Grunz JP. Diffusion-weighted Breast MRI at 3 Tesla: Improved Lesion Visibility and Image Quality with a Combination of Water-excitation and Spectral Fat Saturation. Acad Radiol 2023;30:1773-83. [Crossref] [PubMed]
  23. Taouli B, Koh DM. Diffusion-weighted MR imaging of the liver. Radiology 2010;254:47-66. [Crossref] [PubMed]
  24. Nasu K, Kuroki Y, Nawano S, Kuroki S, Tsukamoto T, Yamamoto S, Motoori K, Ueda T. Hepatic metastases: diffusion-weighted sensitivity-encoding versus SPIO-enhanced MR imaging. Radiology 2006;239:122-30. [Crossref] [PubMed]
  25. Park MY, Byun JY. Understanding the mathematics involved in calculating apparent diffusion coefficient maps. AJR Am J Roentgenol 2012;199:W784. [Crossref] [PubMed]
  26. Tavakoli A, Attenberger UI, Budjan J, Stemmer A, Nickel D, Kannengiesser S, Morelli JN, Schoenberg SO, Riffel P. Improved Liver Diffusion-Weighted Imaging at 3 T Using Respiratory Triggering in Combination With Simultaneous Multislice Acceleration. Invest Radiol 2019;54:744-51. [Crossref] [PubMed]
  27. Kandpal H, Sharma R, Madhusudhan KS, Kapoor KS. Respiratory-triggered versus breath-hold diffusion-weighted MRI of liver lesions: comparison of image quality and apparent diffusion coefficient values. AJR Am J Roentgenol 2009;192:915-22. [Crossref] [PubMed]
  28. Kwee TC, Takahara T, Koh DM, Nievelstein RA, Luijten PR. Comparison and reproducibility of ADC measurements in breathhold, respiratory triggered, and free-breathing diffusion-weighted MR imaging of the liver. J Magn Reson Imaging 2008;28:1141-8. [Crossref] [PubMed]
  29. Ivancevic MK, Kwee TC, Takahara T, Ogino T, Hussain HK, Liu PS, Chenevert TL. Diffusion-weighted MR imaging of the liver at 3.0 Tesla using TRacking Only Navigator echo (TRON): a feasibility study. J Magn Reson Imaging 2009;30:1027-33. [Crossref] [PubMed]
  30. Saake M, Seuß H, Riexinger A, Bickelhaupt S, Hammon M, Uder M, Laun FB. Image Quality and Detection of Small Focal Liver Lesions in Diffusion-Weighted Imaging: Comparison of Navigator Tracking and Free-Breathing Acquisition. Invest Radiol 2021;56:579-90. [Crossref] [PubMed]
  31. Li XM, Ma FZ, Quan XY, Zhang XC, Xiao BH, Wáng YXJ. Repeatability and reproducibility comparisons of liver IVIM imaging with free-breathing or respiratory-triggered sequences. NMR Biomed 2024;37:e5080. [Crossref] [PubMed]
  32. Wang Y, Qi H, Wang Y, Xiao M, Xiang C, Dong J, Chen H. Free-breathing simultaneous water-fat separation and T1 mapping of the whole liver (SWALI) with isotropic resolution using 3D golden-angle radial trajectory. Quant Imaging Med Surg 2023;13:912-23. [Crossref] [PubMed]
  33. Filipe JP, Curvo-Semedo L, Casalta-Lopes J, Marques MC, Caseiro-Alves F. Diffusion-weighted imaging of the liver: usefulness of ADC values in the differential diagnosis of focal lesions and effect of ROI methods on ADC measurements. MAGMA 2013;26:303-12. [Crossref] [PubMed]
  34. Bruegel M, Holzapfel K, Gaa J, Woertler K, Waldt S, Kiefer B, Stemmer A, Ganter C, Rummeny EJ. Characterization of focal liver lesions by ADC measurements using a respiratory triggered diffusion-weighted single-shot echo-planar MR imaging technique. Eur Radiol 2008;18:477-85. [Crossref] [PubMed]
  35. Kwee TC, Takahara T, Niwa T, Ivancevic MK, Herigault G, Van Cauteren M, Luijten PR. Influence of cardiac motion on diffusion-weighted magnetic resonance imaging of the liver. MAGMA 2009;22:319-25. [Crossref] [PubMed]
  36. Mürtz P, Flacke S, Träber F, van den Brink JS, Gieseke J, Schild HH. Abdomen: diffusion-weighted MR imaging with pulse-triggered single-shot sequences. Radiology 2002;224:258-64. [Crossref] [PubMed]
  37. Dietrich O, Biffar A, Baur-Melnyk A, Reiser MF. Technical aspects of MR diffusion imaging of the body. Eur J Radiol 2010;76:314-22. [Crossref] [PubMed]
Cite this article as: Chen Z, Xing Z, Zheng E, Luo M, Fu C, Li G, Benkert T, Xue Y, Sun B. Optimization of motion-corrected liver diffusion-weighted imaging at 3 Tesla (3T): incorporating complex averaging and reparametrized sinc fatsat pulse with optimized water excitation pulse. Quant Imaging Med Surg 2024;14(9):6579-6589. doi: 10.21037/qims-24-340

Download Citation