Adiabatically prepared spin-lock could reduce the R1ρ dispersion
Original Article

Adiabatically prepared spin-lock could reduce the R dispersion

Ping Wang^

Neuroimaging Innovation Center, Barrow Neurological Institute, Phoenix, AZ, USA

^ORCID: 0000-0003-0490-005X.

Correspondence to: Ping Wang, PhD. Neuroimaging Innovation Center, Barrow Neurological Institute, 350 W Thomas Rd., Phoenix, AZ 85013, USA. Email: Ping.Wang@BarrowNeuro.org.

Background: R (or spin-lock) imaging is prone to artifacts arising from field inhomogeneities that may impact the R quantification. Previous research has proposed two types of method to manage the artifacts in continuous-wave constant amplitude spin-lock, one is based on the composite block pulses to compensate for the field imperfections, another category uses adiabatic pulses in the R pre-pulse to excite and reverse the magnetization (named adiabatic prepared approach). Although both methods have proved their efficiency in alleviating artifacts, we observed that the adiabatic pulse approach could produce much lower R dispersion in human knee cartilage than the block pulse method (characterized by the R difference ∆R =11.4 Hz (from spin-lock field 50 to 500 Hz) for the block pulse method vs. ∆R =4.5 Hz for the adiabatic pulse approach). Prompted by this observation, the purpose of this study was to investigate the underlying factors that may affect the R dispersion through numerical simulations based on the two-pool exchanging Bloch-McConnell equations.

Methods: The effects of free water pool size Pa (from 0.80 to 0.95), chemical exchange rate kb (from the bound to free water pool, ranged from 500 to 3,000 Hz), adiabatic pulse duration Tp (from 5.0 to 25 ms), and the chemical shift of the bound pool ppmb (from 1.0 to 5.0 ppm) were examined on the degree of the R dispersion for the two R imaging methods.

Results: In general, the greater the ppmb, kb, Tp, and the smaller Pa, the more significant difference in R dispersion between the block and adiabatic approaches, with the dispersion curve of the adiabatic method becoming flatter.

Conclusions: The adiabatic prepared approach may compromise the R dispersion, the effect is determined by the combination of the tissue and radiofrequency (RF) pulse properties. It is suggested that care should be taken when using the adiabatically prepared approach to study R dispersion.

Keywords: R1ρ dispersion; chemical exchange; adiabatic pulse; Bloch-McConnell equations


Submitted Sep 30, 2021. Accepted for publication Nov 07, 2022. Published online Dec 09 2022.

doi: 10.21037/qims-21-959


Introduction

R (=1/T), the spin lattice relaxation rate in the rotating frame, has been used extensively to probe the relatively slow macromolecular processes, making it a practical tool for gaining information about water spin dynamics and interactions with endogenous macromolecules (1). Depending on the tissue types and the changes in tissue component and microenvironment, T value may increase or decrease in diseases such as osteoarthritis (2-5), intervertebral disc degeneration (6,7), fibrosis (8,9), and liver steatosis (10,11). R imaging involves the application of specific radiofrequency (RF) fields (called spin-lock fields) that can influence the R relaxation processes so that the R value varies with the strength of the RF pulse used. This dispersion of relaxation rate R with the spin-lock field may be used to quantify the dynamic properties in biological tissues. There are different mechanisms potentially contributing to the R relaxation, i.e., dipolar-dipolar interaction, diffusion, and chemical exchange (12-16). However, literature regarding R relaxation mechanisms at 3T is somewhat inconsistent, with some groups reporting that similar to T2, the dominating factor in T relaxation is dipolar interaction (5,15,17,18), contrary to another study by Li et al. (19) where only a minor magic angle effect (associated with dipolar interaction) in cadaveric human femoral-tibial cartilage was observed. In addition, some studies reported that chemical exchange may be a main contributor to T relaxation at high static fields (3T and above) and which leads to a significant T dispersion in certain tissues (20-23). Based on the previous results (24-26), it was suggested that R at very low locking fields (≤200 Hz) may reflect diffusion of tissue water molecules within field gradients caused by local magnetic field inhomogeneities, however, at higher locking fields, chemical exchange effects may dominate (14,27). Because the time scales of these two effects are so different, these two processes are readily separated (14). The dispersion of R has been used to assess the vascular properties of muscles (14) and the water diffusion through susceptibility gradient in tumors (27); this was also used to characterize the contribution of chemical exchange from macromolecules that consist of labile protons (associated with hydroxyls, amides, amines) exchanging at an appropriate rate with the tissue water (20,25).

One of the challenges in R imaging is that it is prone to artifacts arising from field inhomogeneities, which may greatly impact the R quantification accuracy if not corrected. Previous studies have addressed the issues using different approaches, the commonly used approach is based on a composite block RF pulse clusters combining the rotary echo method (28) with a 180-degree refocusing pulse to compensate for the field imperfections (29), however the performance of the method relies on the perfection of the 180-degree pulse; another approach uses adiabatic pulses in the R pre-pulse to tip down and back the magnetizations (30,31), which is termed “adiabatic prepared approach” in this paper to distinguish from the chains of adiabatic pulses used in (32,33). While both methods work well in terms of mitigating image artifacts, their influence on the dispersion degree has not been studied previously, with only one group showing that different image contrasts can be achieved by manipulating the pulse properties of the adiabatic and continuous-wave (CW) constant amplitude R imaging experiments (32,33).

In our recent R imaging in human knee cartilage, we observed that the degree of R dispersion using the adiabatic approach (hyperbolic secant, HS1) was significantly lower than the block pulse approach. Prompted by this observation, we investigate whether and how the properties of tissue and RF pulses may influence the R dispersion through numerical simulations. Although the current research progress on the origin of R relaxation remain inconsistent, in this study we considered only the chemical exchange effect as its contribution appears to increase with the increasing static field as well as the locking field (20-22). We employed a two-pool model (the bulk water pool “a” and exchangeable solute pool “b”) and tracked the magnetization during the whole R pre-pulse by solving the numerical solutions of the Bloch-McConnell equations (34). The R values were computed exactly following the data fitting procedure in real applications, and the R dispersion curves were extracted by plotting the R values vs. spin-lock strengths. Specially, the simulations examined the dependencies of R dispersion on (I) water pool fractional size Pa, (II) exchange rate from the solute to water pools, kb, (III) the duration of the adiabatic pulse Tp, and (IV) chemical shift of the solute pool, ppmb. We found that in general, with the increase of chemical exchange rate kb, adiabatic pulse duration Tp, and chemical shift ppmb, as well as the decrease of water pool size ratio Pa, the difference in R dispersion between the block and adiabatic methods increasingly differs, with the R dispersion curve of the adiabatic method becoming flatter. We present the following article in accordance with the MDAR reporting checklist (available at https://qims.amegroups.com/article/view/10.21037/qims-21-959/rc).


Methods

This study was conducted in accordance with the Declaration of Helsinki (as revised in 2013) and was approved by the local IRB (Institutional Review Board) and written informed consent was obtained from all participants.

Adiabatic pulse

Adiabatic pulse is both amplitude and frequency modulated. For the most prevalent hyperbolic secant (HS) family, the amplitude and frequency modulations have following forms:

w1(t)=w1maxsech((βtTp)n)

Δw(t)=wRF(t)w0=Asech2((βτTp)n)dτ

where w1max is the maximum value of w1(t), w0 is the on-resonance frequency, wRF is the carrier frequency of the pulse, Tp is the pulse duration, and A determines the amplitude of the frequency sweep, and β is a dimensionless truncation factor. One fundamental property of the adiabatic pulse is the time-bandwidth product given by:

TBW=TpBW

where BW is the bandwidth of the pulse. With these definitions, the frequency sweep amplitude

A=πBW=πTBW/Tp

When n=1, Eqs. [1] and [2] are simplified to the HS1 pulse:

w1(t)=w1maxsech(βTpt)

Δw(t)=wRF(t)w0=Asech2(βτTp)dτ=Atanh(βTpt)

The adiabatic condition is that the direction of effective magnetic field does not change much during one period of precession of the magnetization about the effective field. Under this condition, the adiabatic full passage pulse (AFP, one cycle of the hyperbolic secant function) is able to nutate the magnetization 180°, for instance, from z axis to -z axis despite of the RF inhomogeneity. One property of the adiabatic pulse is that the adiabatic half passage (AHP, half duration of the AFP) can turn the magnetization 90°.

Field inhomogeneities insensitive R pre-pulse

In the typical spin-lock experiment, the equilibrium magnetization is nutated to the transverse plane by an RF pulse, the magnetization is then spin-locked by a continuous-wave constant amplitude spin-lock RF pulse for a period to generate the R contrast, finally the magnetization is tipped back to the z-direction followed by signal acquisition. Figure 1 shows the R imaging pulses that are commonly used to overcome artifacts from field inhomogeneities, details can be found in (29-31).

Figure 1 Diagram of the continuous-wave constant amplitude spin-lock pulse cluster to overcome field inhomogeneities. (A) Composite block R pre-pulse. Magnetization is tipped a flip angle αx about x axis by the first pulse, the spin-lock pulse (green box) is separated by a 180° refocusing pulse applied about the y axis at the middle of the spin-lock, forming two spin-lock segments with opposite phases (+y and -y). Finally, the R prepared magnetization is turned back to -z axis by the second αx pulse. (B) An AHP sequence is applied at the front of the R pre-pulse to tip the magnetization a 90º, with the AM and FM derived from the HS1 pulse. The spin-lock pulse is applied with a duration of TSL and amplitude of FSL, at the end the rAHP pulse tips the magnetization back to the z axis. RF, radiofrequency; TSL, spin-lock time; AM, amplitude modulation; FM, frequency modulation; AHP, adiabatic half passage; HS1, hyperbolic secant (n=1); FSL, spin-lock frequency; rAHP, reverse AHP.

Previous observation

Previous study used R dispersion to assess chemical exchange in knee cartilage with the R pulse as in Figure 1A, significant dispersion was observed for a spin-lock frequency range from 0 to 550 Hz on a Philips 3T Achieva scanner (Philips Healthcare, Cleveland, OH, USA) (20). However, in our later experiments with the adiabatic approach as in Figure 1B on healthy volunteers (n=3), only negligible dispersion was found, see Figure 2, which prompts us to study whether the properties of the tissue and pulse affect the dispersion degree.

Figure 2 R imaging in human knee cartilage. (A) R map in knee cartilage at different locking fields for block (upper row) and adiabatic (lower row) pulses as in Figure 1. Left to right corresponds to the FSL = [50, 100, 200, 300, 400, 500] Hz. For the block pulse approach, TSL = [2, 22, 42, 62, 82] ms; for the adiabatic pulse approach, TSL = [1, 21, 41, 61, 81] ms. Other parameters: FOV =148×128 mm2, matrix size =320×320, thickness =4 mm. Data were acquired by the Turbo Spin Echo sequence. (B) R dispersion from the knee cartilage, median values of R were used in the plot. FSL, spin-lock frequency; TSL, spin-lock time; FOV, field of view.

Numerical simulations

We only investigate the chemical exchange effect on R relaxation in the simulations. We examine whether the dispersion of the block and adiabatic methods behaves differently to the properties of tissue and the RF at the main magnetic field of 3.0T. We assume a two-pool exchange tissue model, i.e., the bulk water pool “a” and a smaller metabolite pool “b” with certain chemical shift and exchange rate. The behavior of this two-pool exchanging system may be analyzed using the Bloch-McConnell equations as below (34):

ddt=(MxaMxbMyaMybMzaMzb)=(R2akakbΔw0aw1z0w1y0kaR2bkb0Δw0bw1z0w1yΔw2aw1z0R2akakbw1x00Δw0bw1zkaR2bkb0w1xw1y0w1x0R1akakb0w1y0w1xkaR1bkb)(MxaMxbMyaMybMzaMzb)+(0000R1aM0aR1bM0b)

where Mx,y,za,b denotes the magnetization along x (or y, z) axis for pool a and pool b, M0a,b is the equilibrium magnetization of pool a or b, R1,2a,b the longitudinal (or transverse) relaxation rate for the two pools; ka and kb are chemical exchange rates from pool a to pool b, and from pool b to pool a respectively. In addition, w1x,1y,1z describes the applied RF field in x (or y, z) axis respectively, and Δw0a,b is the chemical shift terms for pool a or b. Finally, the pool fractional size is defined as Pa and Pb, with Pa + Pb =1, and also ka = (kb · Pb)/Pa.

The numerical solutions of the Bloch-McConnell equations were obtained using Matlab (Mathworks, R2018a) codes by solving the ordinary differential equations, the magnetizations were tracked from excitation to reversion during the R pre-pulse, exactly following the real R experiments. The MRI signals (the final magnetizations that have been turned to z axis) were fitted to a two-parameter mono-exponential model

S=S0exp(TSLR1ρ)

We mainly focus on whether the tissue and RF properties affect the dispersion for the block and adiabatic methods, so the field inhomogeneities were not considered in the simulations. An AHP HS1 pulse (Eqs. [5] and [6]) was selected for the adiabatic method, with β =4.0, TBW =10, and w1max=2πγB1max (γ is the gyromagnetic ratio 42.58×106 Hz/Tesla, and B1max=15.0μT is the maximum B1 assumed at 3T), the frequency sweep amplitude A can be derived from Eq. [4]. The experiments were performed with a series of TSLs {=[0, 20, 40, 60] ms} and FSLs from 50 to 1,000 Hz with an increment of 25 Hz. The dispersion curves of both methods were compared under varied values of Pa, kb, Tp (AHP HS1 pulse duration), and ppmb (pool b chemical shift). Since the block pulse duration is almost always chosen for the shortest value (<2 ms), its effect was not considered in the simulations.

To investigate how the parameters (Pa, kb, Tp, ppmb) affect R dispersion, four situations were considered and for each scenario one of the four parameters was treated as a variable within a certain range while the other three parameters remained fixed but with two options: a low and a high value, which led to eight subcases for each situation. Specifically, the simulations were organized as below: (I) Pa varies from 0.80 to 0.95, kb =2,000 or 500 Hz, Tp =10 or 25 ms, and ppmb =1.0 or 5.0 ppm; (II) kb varies from 500 to 3,000 Hz, Pa =0.95 or 0.80, Tp =10 or 25 ms, and ppmb =1.0 or 5.0 ppm; (III) Tp varies from 5.0 to 25 ms, Pa =0.95 or 0.80, kb =2,000 or 500 Hz, and ppmb =1.0 or 5.0 ppm; (IV) ppmb varies from 1.0 to 5.0 ppm, Pa =0.95 or 0.80, kb =2,000 or 500 Hz, and Tp =10 or 25 ms.


Results

Figure 3 shows the comparisons of R dispersion between the block and adiabatic methods, for a series of Pa (0.80 to 0.95 in 0.05 increments) and various combinations of kb (500 vs. 2,000 Hz), Tp (10 vs. 25 ms), and ppmb (1.0 vs. 5.0 ppm). It is seen that for different kb and Tp, the dispersion curves between the block and AHP methods largely match well for the range of Pa at the small ppmb of 1.0 (Figure 3A,3C,3E,3G), though Tp =25 ms appears to have greater error (Figure 3C,3G). However, with a large ppmb = 5.0, the dispersion curves increasingly differ with the decrease of Pa (Figure 3B,3D,3F,3H), the difference becomes more evident at large Tp than at short Tp (Figure 3D vs. Figure 3B, and Figure 3H vs. Figure 3F). Also, the dispersion curves of the adiabatic method appear much flatter at large ppmb and Tp (Figure 3D).

Figure 3 Comparison of R dispersion using adiabatic pulses vs. block pulses for excitation and reversion. The simulation parameters are T1a =4,000 ms, T2a =2,000 ms, T1b =1,250 ms. T2b =35 ms, TSL = [0, 20, 40, 60] ms, FSL = 0 to 1,000 Hz with an increment of 25 Hz. The parameters kb, Tp, and ppmb are specified in (A) – (H), where the comparison is performed for a range of Pa values [0.80, 0.85, 0.90, 0.95]. The curved arrow in (A) shows the direction of Pa increase, which is applicable to other subfigures (B) – (H). TSL, spin-lock time; FSL, spin-lock frequency.

Figure 4 shows the comparisons of R dispersion for a range of kb (500 to 3,000 Hz in 500 increments), with the combination of Pa (0.80 vs. 0.95), Tp (10 vs. 25 ms), and ppmb (1.0 vs. 5.0). Generally, the dispersion curves of the two methods are well close at the small ppmb = 1.0 (Figure 4A,4C,4E,4G). At the large ppmb =5.0, a small Pa (=0.80) can cause a large difference between the dispersion curves for both the pulse durations (Figure 4F,4H).

Figure 4 Comparison of R dispersion using adiabatic pulses vs. block pulses for excitation and reversion. The simulation parameters are T1a =4,000 ms, T2a =2,000 ms, T1b =1,250 ms. T2b =35 ms, TSL = [0, 20, 40, 60] ms, FSL = 0 to 1,000 Hz with an increment of 25 Hz. The parameters Pa, Tp, and ppmb are specified in (A) – (H), where the comparison is performed for a range of kb values [500, 1,000, 1,500, 2,000, 2,500, 3,000] Hz. The curved arrow in (A) shows the direction of kb increase, which is applicable to other subfigures (B) – (H). TSL, spin-lock time; FSL, spin-lock frequency.

Figure 5 demonstrates the comparisons with a range of Tp (5.0 to 25 ms in 5.0 increments) under different combinations of Pa (0.80 vs. 0.95), kb (500 vs. 3,000 Hz), and ppmb (1.0 vs. 5.0). For all situations, the dispersion curves between the two methods match very well except for the case with Pa =0.80, kb=2,000 Hz, and ppmb =5.0, see Figure 5F, where the difference increases with the pulse duration Tp.

Figure 5 Comparison of R dispersion using adiabatic pulses vs. block pulses for excitation and reversion. The simulation parameters are T1a =4,000 ms, T2a =2,000 ms, T1b =1,250 ms. T2b =35 ms, TSL = [0, 20, 40, 60] ms, FSL =0 to 1,000 Hz with an increment of 25 Hz. The parameters Pa, kb, and ppmb are specified in (A) – (H), where the comparison is performed for a range of Tp values [5, 10, 15, 20, 25] ms. The curved arrow in (A) shows the direction of Tp increase, which is applicable to other subfigures (B) – (H). For the block pulse method, the shortest pulse duration is always used so there is only one curve for the block pulse method in each of the subfigure. TSL, spin-lock time; FSL, spin-lock frequency.

Finally, Figure 6 shows the situation for a range of ppmb (1.0 to 5.0 in 1.0 increment) and different combinations of Pa (0.80 vs. 0.95), kb (500 vs. 2,000 Hz), and Tp (10 vs. 25 ms). It shows that at small Pa (=0.80), the dispersion difference between the two methods tends to increase but a large kb (=2,000 Hz) and large ppmb (=5.0) exacerbate the difference, see Figure 6E,6F, in which the dispersion of the adiabatic approach appears flatter.

Figure 6 Comparison of R dispersion using adiabatic pulses vs. block pulses for excitation and reversion. The simulation parameters are T1a =4,000 ms, T2a =2,000 ms, T1b =1,250 ms. T2b =35 ms, TSL = [0, 20, 40, 60] ms, FSL =0 to 1,000 Hz with an increment of 25 Hz. The parameters Pa, kb, and Tp are specified in (A) – (H), where the comparison is performed for a range of ppmb values [1.0, 2.0, 3.0, 4.0, 5.0]. The curved arrow in (A) shows the direction of ppmb increase, which is applicable to other subfigures (B) – (H). TSL, spin-lock time; FSL, spin-lock frequency.

Discussion

R dispersion holds great potential to assess molecular dynamics in biological tissues and has been exploited as an important method for the early diagnosis of diseases. Although other mechanisms such as dipolar effect and diffusion may contribute, previous studies reported that chemical exchange may be a main contributor to the R dispersion at higher magnetic fields (14,15,20,21). Exploiting the R dispersion holds great potential for the characterization of tissue composition and the physicochemical changes associated with pathology.

In this study, we investigated the difference of R dispersion between the block and adiabatic methods under solely the effect of chemical exchange at 3T using a two-pool model system. We considered the influence of different parameters of pool size Pa, chemical exchange rate kb, adiabatic pulse duration Tp, and the chemical shift of the solute pool ppmb. It is seen that generally, the larger ppmb, kb, and Tp, and the smaller Pa, the more pronounced difference in the R dispersion between the block and adiabatic methods, with the dispersion curve of the adiabatic method appears flatter. Although the dispersion difference is determined by the combination of these parameters, it seems that ppmb is a more sensitive factor. In contrast, the smaller ppmb, kb, and Tp, and the greater Pa, the better coincidence between the two methods.

This study has several limitations. First, wider ranges of parameter were used in the simulations to better reveal the effect of the parameter on the R dispersion, so the dispersion curves in some scenarios might look different from those observed in biological tissues (in scale and/or pattern). Second, only chemical exchange was considered and the readout sequence was neglected. In the real R experiment in biological tissues, however, other relaxation channels such as diffusion, dipolar-dipolar interaction may also contribute to the R dispersion. Also depending on the readout sequence, the R contrast may be compromised by the relaxation parameters and pulse sequence parameters (35). Third, this study was based on simulations only, however, systematically designed real experiments would be essential in the validation of the theoretical analysis and simulations. Our future work will focus more on the investigation of R relaxation mechanisms and their validation. Nevertheless, the simulations showed that at certain conditions, the adiabatic pulse method may lead to significantly lower R dispersion than the block pulse approach, as we have observed in knee cartilage imaging (Figure 2), suggesting that care should be taken when using adiabatic approach to study the R dispersion. The difference in R dispersion between the two approaches may be because the tipping pulses (adiabatic vs. block) have different effect on the magnetizations, and the effect depends on the properties of both the tipping pulses and tissues.

Previous R studies in cartilage mainly focused on whether R values are relevant to the concentration of glycosaminoglycan (GAG), a side chain of proteoglycan and clinically an indication of osteoarthritis, although there have been conflicting conclusions regarding the origins of the R contrast (3,5,36-38). There were very few studies investigating the effect of both proteoglycan and collagen on R values in cartilage specimens, with one study concluding that degradation of proteoglycans and collagen fibers in the articular cartilage increased the articular cartilage T (=1/R) value (39), and another study being that T may be primarily determined by collagen concentration but the molecular level interactions associated with collagen/GAG may be contributing in an important way to T (4). In our early study of R dispersion in knee cartilage with the block pulse approach (20), we speculated that the chemical exchange was mainly between free water and hydroxyls in GAGs. Since the chemical shift of hydroxyls is small about 1.0–1.2 ppm, it may not account for the difference of the R curves for the block and adiabatic methods (Figure 2). According to the simulations in our paper, the important reason causing the dispersion difference would be the large chemical shift ppmb, which perhaps suggests that other substance in the cartilage, for instance the most abundant collagens, may also participate in the chemical exchange process, as the collagen macromolecules have exchangeable amine and amide protons (40) with large chemical shifts (amide NH: 5.5–8 ppm, and amine NH2: 0.5–3.0 ppm) (41). This interpretation differs from some previous conclusions about the chemical exchange in knee cartilage.

We observed a “dip” sometimes occurring at the lower FSLs (for instance <100 Hz) in Figures 3-6, which we have also observed sometimes in the real R experiments in biological tissues. We interpret this as a situation where the spins are not “locked” efficiently about the locking field direction at lower locking fields, the “dip” is generally less prominent at large water pool size Pa, small exchange rate kb, and small chemical shift of the bound pool ppmb.


Conclusions

In conclusion, it is suggested that care should be taken when using the adiabatically prepared approach to study R dispersion. The adiabatic approach may compromise the R dispersion, the effect is determined by the combination of the tissue and RF properties.


Acknowledgments

Funding: This work was supported by the Barrow Neurological Foundation (No. 455003033568).


Footnote

Reporting Checklist: The author has completed the MDAR reporting checklist. Available at https://qims.amegroups.com/article/view/10.21037/qims-21-959/rc

Conflicts of Interest: The author has completed the ICMJE uniform disclosure form (available at https://qims.amegroups.com/article/view/10.21037/qims-21-959/coif). The author has no conflicts of interest to declare.

Ethical Statement: The author is accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013). The study was approved by the local IRB (Institutional Review Board) and written informed consent was obtained from all participants.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Liepinsh E, Otting G. Proton exchange rates from amino acid side chains--implications for image contrast. Magn Reson Med 1996;35:30-42. [Crossref] [PubMed]
  2. Borthakur A, Mellon E, Niyogi S, Witschey W, Kneeland JB, Reddy R. Sodium and T1rho MRI for molecular and diagnostic imaging of articular cartilage. NMR Biomed 2006;19:781-821. [Crossref] [PubMed]
  3. Li X, Benjamin Ma C, Link TM, Castillo DD, Blumenkrantz G, Lozano J, Carballido-Gamio J, Ries M, Majumdar S. In vivo T(1rho) and T(2) mapping of articular cartilage in osteoarthritis of the knee using 3 T MRI. Osteoarthritis Cartilage 2007;15:789-97. [Crossref] [PubMed]
  4. Menezes NM, Gray ML, Hartke JR, Burstein D. T2 and T1rho MRI in articular cartilage systems. Magn Reson Med 2004;51:503-9. [Crossref] [PubMed]
  5. Mlynárik V, Szomolányi P, Toffanin R, Vittur F, Trattnig S. Transverse relaxation mechanisms in articular cartilage. J Magn Reson 2004;169:300-7. [Crossref] [PubMed]
  6. Paul CPL, Smit TH, de Graaf M, Holewijn RM, Bisschop A, van de Ven PM, Mullender MG, Helder MN, Strijkers GJ. Quantitative MRI in early intervertebral disc degeneration: T1rho correlates better than T2 and ADC with biomechanics, histology and matrix content. PLoS One 2018;13:e0191442. [Crossref] [PubMed]
  7. Wáng YX, Zhang Q, Li X, Chen W, Ahuja A, Yuan J. T1ρ magnetic resonance: basic physics principles and applications in knee and intervertebral disc imaging. Quant Imaging Med Surg 2015;5:858-85. [PubMed]
  8. Hectors SJ. Is MRI relaxometry parameter T1ρ specific to fibrosis or confounded by concomitant pathological features? Quant Imaging Med Surg 2020;10:2408-10. [Crossref] [PubMed]
  9. Wang F, Colvin DC, Wang S, Li H, Zu Z, Harris RC, Zhang MZ, Gore JC. Spin-lock relaxation rate dispersion reveals spatiotemporal changes associated with tubulointerstitial fibrosis in murine kidney. Magn Reson Med 2020;84:2074-87. [Crossref] [PubMed]
  10. Zhao F, Zhou N, Wang X, Wang JL, Zhong WX, Deng M, Zheng CJ, He J, Yan SX, Wáng YXJ. T1rho shortening effect of fat in liver steatosis after fat suppression: approximate estimation in a methionine and choline-deficient (MCD) diet rat model. Quant Imaging Med Surg 2021;11:870-5. [Crossref] [PubMed]
  11. Okuaki T, Takayama Y, Nishie A, Ogino T, Obara M, Honda H, Miyati T, Van Cauteren M. T1ρ mapping improvement using stretched-type adiabatic locking pulses for assessment of human liver function at 3T. Magn Reson Imaging 2017;40:17-23. [Crossref] [PubMed]
  12. Akella SV, Regatte RR, Wheaton AJ, Borthakur A, Reddy R. Reduction of residual dipolar interaction in cartilage by spin-lock technique. Magn Reson Med 2004;52:1103-9. [Crossref] [PubMed]
  13. Cobb JG, Xie J, Gore JC. Contributions of chemical and diffusive exchange to T1ρ dispersion. Magn Reson Med 2013;69:1357-66. [Crossref] [PubMed]
  14. Adelnia F, Zu Z, Spear JT, Wang F, Harkins KD, Gore JC. Tissue characterization using R1rho dispersion imaging at low locking fields. Magn Reson Imaging 2021;84:1-11. [Crossref] [PubMed]
  15. Shao H, Pauli C, Li S, Ma Y, Tadros AS, Kavanaugh A, Chang EY, Tang G, Du J. Magic angle effect plays a major role in both T1rho and T2 relaxation in articular cartilage. Osteoarthritis Cartilage 2017;25:2022-30. [Crossref] [PubMed]
  16. Pang Y, Palmieri-Smith RM, Malyarenko DI, Swanson SD, Chenevert TL. A unique anisotropic R2 of collagen degeneration (ARCADE) mapping as an efficient alternative to composite relaxation metric (R2 -R1ρ) in human knee cartilage study. Magn Reson Med 2019;81:3763-74. [Crossref] [PubMed]
  17. Du J, Statum S, Znamirowski R, Bydder GM, Chung CB. Ultrashort TE T1ρ magic angle imaging. Magn Reson Med 2013;69:682-7. [Crossref] [PubMed]
  18. Mlynárik V, Trattnig S, Huber M, Zembsch A, Imhof H. The role of relaxation times in monitoring proteoglycan depletion in articular cartilage. J Magn Reson Imaging 1999;10:497-502. [Crossref] [PubMed]
  19. Li X, Cheng J, Lin K, Saadat E, Bolbos RI, Jobke B, Ries MD, Horvai A, Link TM, Majumdar S. Quantitative MRI using T1ρ and T2 in human osteoarthritic cartilage specimens: correlation with biochemical measurements and histology. Magn Reson Imaging 2011;29:324-34. [Crossref] [PubMed]
  20. Wang P, Block J, Gore JC. Chemical exchange in knee cartilage assessed by R1ρ (1/T1ρ) dispersion at 3T. Magn Reson Imaging 2015;33:38-42. [Crossref] [PubMed]
  21. Spear JT, Gore JC. New insights into rotating frame relaxation at high field. NMR Biomed 2016;29:1258-73. [Crossref] [PubMed]
  22. Cobb JG, Xie J, Gore JC. Contributions of chemical exchange to T1ρ dispersion in a tissue model. Magn Reson Med 2011;66:1563-71. [Crossref] [PubMed]
  23. Chopra S, McClung R, Jordan R. Rotating-frame relaxation rates of solvent molecules in solutions of paramagnetic ions undergoing solvent exchange. J Magn Reson 1984;59:361-72.
  24. Cobb JG, Li K, Xie J, Gochberg DF, Gore JC. Exchange-mediated contrast in CEST and spin-lock imaging. Magn Reson Imaging 2014;32:28-40. [Crossref] [PubMed]
  25. Cobb JG, Xie J, Li K, Gochberg DF, Gore JC. Exchange-mediated contrast agents for spin-lock imaging. Magn Reson Med 2012;67:1427-33. [Crossref] [PubMed]
  26. Spear JT, Zu Z, Gore JC. Dispersion of relaxation rates in the rotating frame under the action of spin-locking pulses and diffusion in inhomogeneous magnetic fields. Magn Reson Med 2014;71:1906-11. [Crossref] [PubMed]
  27. Zu Z, Janve V, Gore JC. Spin-lock imaging of intrinsic susceptibility gradients in tumors. Magn Reson Med 2020;83:1587-95. [Crossref] [PubMed]
  28. Charagundla SR, Borthakur A, Leigh JS, Reddy R. Artifacts in T(1rho)-weighted imaging: correction with a self-compensating spin-locking pulse. J Magn Reson 2003;162:113-21. [Crossref] [PubMed]
  29. Witschey WR 2nd, Borthakur A, Elliott MA, Mellon E, Niyogi S, Wallman DJ, Wang C, Reddy R. Artifacts in T1 rho-weighted imaging: compensation for B(1) and B(0) field imperfections. J Magn Reson 2007;186:75-85. [Crossref] [PubMed]
  30. Chen W. Artifacts correction for T1rho imaging with constant amplitude spin-lock. J Magn Reson 2017;274:13-23. [Crossref] [PubMed]
  31. Schuenke P, Koehler C, Korzowski A, Windschuh J, Bachert P, Ladd ME, Mundiyanapurath S, Paech D, Bickelhaupt S, Bonekamp D, Schlemmer HP, Radbruch A, Zaiss M. Adiabatically prepared spin-lock approach for T1ρ-based dynamic glucose enhanced MRI at ultrahigh fields. Magn Reson Med 2017;78:215-25. [Crossref] [PubMed]
  32. Mangia S, Liimatainen T, Garwood M, Michaeli S. Rotating frame relaxation during adiabatic pulses vs. conventional spin lock: simulations and experimental results at 4 T. Magn Reson Imaging 2009;27:1074-87. [Crossref] [PubMed]
  33. Michaeli S, Sorce DJ, Garwood M. T2rho and T1rho Adiabatic Relaxations and Contrasts. Curr Anal Chem 2008;4:8-25. [Crossref]
  34. McConnell HM. Reaction rates by nuclear magnetic resonance. J Chem Phys 1958;28:430-1. [Crossref]
  35. Chen W. Errors in quantitative T1rho imaging and the correction methods. Quant Imaging Med Surg 2015;5:583-91. [PubMed]
  36. Son M, Goodman SB, Chen W, Hargreaves BA, Gold GE, Levenston ME. Regional variation in T1ρ and T2 times in osteoarthritic human menisci: correlation with mechanical properties and matrix composition. Osteoarthritis Cartilage 2013;21:796-805. [Crossref] [PubMed]
  37. van Tiel J, Kotek G, Reijman M, Bos PK, Bron EE, Klein S, Nasserinejad K, van Osch GJ, Verhaar JA, Krestin GP, Weinans H, Oei EH. Is T1ρ Mapping an Alternative to Delayed Gadolinium-enhanced MR Imaging of Cartilage in the Assessment of Sulphated Glycosaminoglycan Content in Human Osteoarthritic Knees? An in Vivo Validation Study. Radiology 2016;279:523-31. [Crossref] [PubMed]
  38. Regatte RR, Akella SV, Lonner JH, Kneeland JB, Reddy R. T1rho relaxation mapping in human osteoarthritis (OA) cartilage: comparison of T1rho with T2. J Magn Reson Imaging 2006;23:547-53. [Crossref] [PubMed]
  39. Choi WS, Yoo HJ, Hong SH, Choi JY. The Effects of Proteoglycan and Type II Collagen on T1ρ Relaxation Time of Articular Cartilage. J Korean Soc Radiol 2015;72:108-14. [Crossref]
  40. Book. Cartilage Imaging: Significance, Techniques, and New Developments. Springer, 2011.
  41. Webpage. Available online: https://tigerweb.towson.edu/jdiscord/www/paperwork_fall2018/labinfofall2018/spectroscopyvalues.pdf, accessed October 20 2022.
Cite this article as: Wang P. Adiabatically prepared spin-lock could reduce the R dispersion. Quant Imaging Med Surg 2023;13(2):763-775. doi: 10.21037/qims-21-959

Download Citation